Dynamic measurements and uncertainty estimation of clinical thermometers using Monte Carlo method

Clinical thermometers in intensive care units are used for the continuous measurement of body temperature. This study describes a procedure for dynamic measurement uncertainty evaluation in order to examine the requirements for clinical thermometer dynamic properties in standards and recommendations. In this study thermistors were used as temperature sensors, transient temperature measurements were performed in water and air and the measurement data were processed for the investigation of thermometer dynamic properties. The thermometers were mathematically modelled. A Monte Carlo method was implemented for dynamic measurement uncertainty evaluation. The measurement uncertainty was analysed for static and dynamic conditions. Results showed that dynamic uncertainty is much larger than steady-state uncertainty. The results of dynamic uncertainty analysis were applied on an example of clinical measurements and were compared to current requirements in ISO standard for clinical thermometers. It can be concluded that there was no need for dynamic evaluation of clinical thermometers for continuous measurement, while dynamic measurement uncertainty was within the demands of target uncertainty. Whereas in the case of intermittent predictive thermometers, the thermometer dynamic properties had a significant impact on the measurement result. Estimation of dynamic uncertainty is crucial for the assurance of traceable and comparable measurements.

[1]  Franc Cimerman,et al.  Identification of the dynamic properties of temperature-sensors in natural and petroleum gas , 2002 .

[2]  T Togawa,et al.  Body temperature measurement. , 1985, Clinical physics and physiological measurement : an official journal of the Hospital Physicists' Association, Deutsche Gesellschaft fur Medizinische Physik and the European Federation of Organisations for Medical Physics.

[3]  Clemens Elster,et al.  Efficient implementation of a Monte Carlo method for uncertainty evaluation in dynamic measurements , 2012 .

[4]  L. Muller,et al.  Temperature measurement in intensive care patients: comparison of urinary bladder, oesophageal, rectal, axillary, and inguinal methods versus pulmonary artery core method , 2003, Intensive Care Medicine.

[5]  Clemens Elster,et al.  Analysis of dynamic measurements and determination of time-dependent measurement uncertainty using a second-order model , 2007 .

[6]  Elizabeth Bridges,et al.  Noninvasive measurement of body temperature in critically ill patients. , 2009, Critical care nurse.

[7]  Amine Ismail,et al.  A nonparametric model for sensors used in a dynamic context , 2012 .

[8]  L. Michalski,et al.  Temperature Measurement , 1985 .

[9]  D. Marquardt An Algorithm for Least-Squares Estimation of Nonlinear Parameters , 1963 .

[10]  Clemens Elster,et al.  Deconvolution filters for the analysis of dynamic measurement processes: a tutorial , 2010 .

[11]  D. Nierman,et al.  Core temperature measurement in the intensive care unit , 1991, Critical care medicine.

[12]  C. Ament,et al.  Temperature-Dependent Dynamic Behavior of Process Temperature Sensors , 2015 .

[13]  John S. Steinhart,et al.  Calibration curves for thermistors , 1968 .