On the sensitivity of common gamma-index evaluation methods to MLC misalignments in Rapidarc quality assurance.

PURPOSE In this study the effects of small systematic MLC misalignments and gravitational errors on the quality of Rapidarc treatment plan delivery are investigated with respect to verification measurements with two detector arrays and the evaluation of clinical significance of the error-induced deviations. METHODS Five prostate and six head and neck plans were modified by means of three error types: (1) both MLC banks are opened, respectively, in opposing directions, resulting in larger fields; (2) both MLC banks are closed, resulting in smaller fields; and (3) both MLC banks are shifted for lateral gantry angles, respectively, in the same direction to simulate the effects of gravity on the leaves. Measurements were evaluated with respect to a gamma-index of 3%/3 mm and 2%/2 mm. Dose in the modified plans was recalculated and the resulting dose volume histograms for target and critical structures were compared to those of the unaltered plans. RESULTS The smallest introduced leaf position deviations which fail the >90% criterion for a gamma-index of 2%/2 mm are: (1) 1 mm; (2) 0.5 mm for prostate and 1.0 mm for head and neck cases; and (3) 3 mm corresponding to the error types, respectively. These errors would lead to significant changes in mean PTV dose and would not be detected with the more commonly used 3%/3 mm gamma-index criterion. CONCLUSIONS A stricter gamma-index (2%/2 mm) is necessary in order to detect positional errors of the MLC. Nevertheless, the quality assurance procedure of Rapidarc treatment plans must include a thorough examination of where dose discrepancies occur, and professional judgment is needed when interpreting the gamma-index analysis, since even a >90% passing rate using the 2%/2 mm gamma-index criterion does not guarantee the absence of clinically significance dose deviation.

[1]  Guanghua Yan,et al.  On the sensitivity of patient‐specific IMRT QA to MLC positioning errors , 2009, Journal of applied clinical medical physics.

[2]  Aime M. Gloi,et al.  RapidArc quality assurance through MapCHECK , 2011, Journal of applied clinical medical physics.

[3]  Jon J Kruse,et al.  On the insensitivity of single field planar dosimetry to IMRT inaccuracies. , 2010, Medical physics.

[4]  J. Bedford,et al.  Commissioning of volumetric modulated arc therapy (VMAT). , 2009, International journal of radiation oncology, biology, physics.

[5]  E. W. Shrigley Medical Physics , 1944, British medical journal.

[6]  S Gaede,et al.  Development of a novel ArcCHECK™ insert for routine quality assurance of VMAT delivery including dose calculation with inhomogeneities. , 2012, Medical physics.

[7]  D. González-Castaño,et al.  Evaluation of chamber response function influence on IMRT verification using 2D commercial detector arrays , 2012, Physics in medicine and biology.

[8]  David A Jaffray,et al.  Intensity-modulated arc therapy with dynamic multileaf collimation : an alternative to tomotherapy , 2002 .

[9]  H. Vorwerk,et al.  Two years experience with quality assurance protocol for patient related Rapid Arc treatment plan verification using a two dimensional ionization chamber array , 2011, Radiation oncology.

[10]  Volker Steil,et al.  Evaluation of a 2D detector array for patient-specific VMAT QA with different setups. , 2011, Physics in medicine and biology.

[11]  Ying Niu,et al.  Is RapidArc more susceptible to delivery uncertainties than dynamic IMRT? , 2012, Medical physics.

[12]  R. Kollhoff,et al.  Two-dimensional ionization chamber arrays for IMRT plan verification. , 2006, Medical physics.

[13]  Mike Oliver,et al.  Understanding the impact of RapidArc therapy delivery errors for prostate cancer , 2011, Journal of applied clinical medical physics.

[14]  C. Ling,et al.  Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy. , 1998, Medical physics.

[15]  C Cameron,et al.  Sweeping-window arc therapy: an implementation of rotational IMRT with automatic beam-weight calculation , 2005, Physics in medicine and biology.

[16]  Alejandra Rangel,et al.  The sensitivity of patient specific IMRT QC to systematic MLC leaf bank offset errors. , 2010, Medical physics.

[17]  Benjamin E Nelms,et al.  Per-beam, planar IMRT QA passing rates do not predict clinically relevant patient dose errors. , 2011, Medical physics.

[18]  E. Cirino,et al.  Quality assurance methodology for Varian RapidArc treatment plans , 2010, Journal of applied clinical medical physics.

[19]  Steve Webb,et al.  Single-Arc IMRT? , 2009, Physics in medicine and biology.

[20]  T. Pawlicki,et al.  Biological consequences of MLC calibration errors in IMRT delivery and QA. , 2012, Medical physics.

[21]  Vladimir Feygelman,et al.  Evaluation of a 3D diode array dosimeter for helical tomotherapy delivery QA. , 2010, Medical Dosimetry.

[22]  L. Kochian Author to whom correspondence should be addressed , 2006 .

[23]  Cedric X. Yu,et al.  Intensity-modulated arc therapy with dynamic multileaf collimation: an alternative to tomotherapy. , 1995, Physics in medicine and biology.

[24]  Benjamin E Nelms,et al.  Moving from gamma passing rates to patient DVH-based QA metrics in pretreatment dose QA. , 2011, Medical physics.

[25]  J. Lutterbach,et al.  Impact of Gantry Rotation Time on Plan Quality and Dosimetric Verification – Volumetric Modulated Arc Therapy (VMAT) vs. Intensity Modulated Radiotherapy (IMRT) , 2011, Strahlentherapie und Onkologie.

[26]  Uwe Oelfke,et al.  Development of an optimization concept for arc-modulated cone beam therapy. , 2007, Physics in medicine and biology.

[27]  Karl Otto,et al.  Volumetric modulated arc therapy: IMRT in a single gantry arc. , 2007, Medical physics.

[28]  Gabriela Studer,et al.  Dysphagia in head and neck cancer patients following intensity modulated radiotherapy (IMRT) , 2011, Radiation oncology.

[29]  Daniel A Low,et al.  A fourier analysis of the dose grid resolution required for accurate IMRT fluence map optimization. , 2005, Medical physics.

[30]  P. Xia,et al.  Impact of MLC leaf position errors on simple and complex IMRT plans for head and neck cancer , 2008, Physics in medicine and biology.

[31]  Stine Korreman,et al.  Dosimetric verification of RapidArc treatment delivery , 2009, Acta oncologica.

[32]  S. Padmanabhan,et al.  Characterization of responses of 2d array seven29 detector and its combined use with octavius phantom for the patient-specific quality assurance in rapidarc treatment delivery. , 2012, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[33]  K. Bush,et al.  Clinical significance of multi-leaf collimator positional errors for volumetric modulated arc therapy. , 2010, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[34]  Mike Oliver,et al.  Comparing planning time, delivery time and plan quality for IMRT, RapidArc and tomotherapy , 2009, Journal of applied clinical medical physics.

[35]  Geoffrey G. Zhang,et al.  Evaluation of a new VMAT QA device, or the “X” and “O” array geometries , 2011, Journal of applied clinical medical physics.

[36]  D. Huyskens,et al.  On-line quality assurance of rotational radiotherapy treatment delivery by means of a 2D ion chamber array and the Octavius phantom. , 2007, Medical physics.