Acoustic noise‐optimized verse pulses

Variable‐rate selective excitation RF pulses modulate the slice selection gradients during RF transmission, especially to reduce the total RF power. Amplitude‐modulated slice selection gradients can lead to increased gradient noise, in particular in high‐field MRI where variable‐rate selective excitation techniques are often used. In this work, an algorithm is presented that calculates a variable‐rate selective excitation pulse modulation from given RF pulses with constant slice selection gradient. The algorithm avoids the known acoustic resonance frequencies of the gradient system to minimize sound pressure levels. It was tested with four different slice‐selective RF pulse shapes (Sinc, Gaussian, and two Shinnar‐LeRoux). Sound measurements revealed a reduction of the mean sound pressure level by up to 13dB, and simultaneously, the specific absorption rate was reduced by 55%. Magn Reson Med, 2010. © 2010 Wiley‐Liss, Inc.

[1]  P A Bottomley,et al.  Power deposition in whole-body NMR imaging. , 1981, Medical physics.

[2]  A Macovski,et al.  A reduced power selective adiabatic spin‐echo pulse sequence , 1991, Magnetic resonance in medicine.

[3]  Jun Shen,et al.  New FOCI pulses with reduced radiofrequency power requirements , 2004, Journal of magnetic resonance imaging : JMRI.

[4]  B. Hargreaves,et al.  Variable‐rate selective excitation for rapid MRI sequences , 2004, Magnetic resonance in medicine.

[5]  A R Palmer,et al.  Sound‐Level Measurements and Calculations of Safe Noise Dosage During EPI at 3 T , 2000, Journal of magnetic resonance imaging : JMRI.

[6]  W A Edelstein,et al.  Characterization and prediction of gradient acoustic noise in MR imagers , 1997, Magnetic resonance in medicine.

[7]  R. Mallozzi,et al.  Making MRI quieter. , 2002, Magnetic resonance imaging.

[8]  A. Macovski,et al.  Variable-rate selective excitation , 1988 .

[9]  Kawin Setsompop,et al.  High-flip-angle slice-selective parallel RF transmission with 8 channels at 7 T. , 2008, Journal of magnetic resonance.

[10]  E. Diesch,et al.  Silent echo-planar imaging for auditory FMRI , 2008, Magnetic Resonance Materials in Physics, Biology and Medicine.

[11]  P. Bandettini,et al.  Functional MRI of brain activation induced by scanner acoustic noise , 1998, Magnetic resonance in medicine.

[12]  T. Loenneker,et al.  “Silent” MRI with soft gradient pulses , 1999, Magnetic resonance in medicine.

[13]  Yuval Zur,et al.  Design and optimization for variable rate selective excitation using an analytic RF scaling function. , 2007, Journal of magnetic resonance.

[14]  J. Pauly,et al.  Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm [NMR imaging]. , 1991, IEEE transactions on medical imaging.