Accuracy of centre of pressure measurement using a piezoelectric force platform.
暂无分享,去创建一个
OBJECTIVE
To evaluate the accuracy in determining centre of pressure when using a force platform for stabilometry research.
DESIGN
Technical report.
BACKGROUND
Previous studies have determined the accuracy of locating the centre of pressure by applying force through a single point. In human stabilometry research however, force is distributed across two feet and the centre of pressure located between the feet.
METHODS
Forces were applied to a Kistler force platform using a force applied through a single point as well as when force was distributed between two metal blocks.
RESULTS
Errors in calculating centre of pressure were substantially larger when force was applied through a single point than when the force was distributed between two blocks.
CONCLUSIONS
Evaluating centre of pressure accuracy by applying force through a single point overestimates the errors inherent in human stabilometry research.
RELEVANCE
Evaluating changes in the centre of pressure of human subjects standing on a force platform is a common tool for evaluating control mechanisms of balance. Accurate information regarding the centre of pressure is essential for the validity of this research.
[1] M. Bobbert,et al. Accuracy of determining the point of force application with piezoelectric force plates. , 1990, Journal of biomechanics.