Clinical performance of a system for semiautomated chromosome analysis.

Until recently equipment for automated chromosome analysis has not been used for routine purposes in clinical cytogenetic laboratories. During a 3 1/2-year period the chromosome laboratory of Rigshospitalet has tested the Magiscan chromosome system under routine conditions and performed the first evaluation of its clinical performance. The system consists of an image processor with a light pen for manual interaction connected to a hard-copy printer and a microscope with a TV camera and a motorized scanning stage for eight slides. Automated metaphase finding takes place without operator assistance. An operator is involved in the analysis after the metaphases are located. Using two of these complete systems, we have performed a total of 4,691 chromosome analyses comprising a count of 10 metaphases, of which three were "eyeball" karyotyped and one was "machine" karyotyped. Presently, two-thirds of our prenatal analyses (amniotic-cell cultures) are carried out with these two machines. A third Magiscan system without scanning stage is used as a "karyotyping-only" system to produce hard-copy karyograms in those cases in which metaphases are manually located and counted in the microscope. Since the end of 1984, 4,773 additional machine karyograms have been produced with this system. With a complete system, a prenatal analysis can be carried out in an average of 35 min. The average time for a machine karyotype is 7 min. Since 1984 the productivity of the laboratory has increased 17%-20% without enlarging the staff.