High peak SAR exposure unit with tight exposure and environmental control for in vitro experiments at 1800 MHz

The objective of this study was to develop, optimize, and characterize a flexible and reliable unit for standardized and well-controlled electromagnetic-field exposure of cells in vitro. The technical requirements were high peak and time-averaged exposure of the cells with a minimal temperature rise for the cell cultures, flexible modulation schemes, high uniformity, and low variability of exposure, as well as support of blinded protocols. The developed setup is based on two R18 waveguides resonant at 1800 MHz and operated with a computer-controlled signal and monitoring unit. The cells can be exposed in 35-mm Petri dishes either cultivated as monolayers or in suspension. For cell monolayers, the system provides an efficiency for the specific absorption rate (SAR) of >50 (W/kg)/W input power, a nonuniformity of the SAR distribution of <30%, SAR variability of <6%, and a temperature rise of <0.03/spl deg/C/(W/kg) average SAR. For cell suspensions and provided that the cells are not in the meniscus area, a SAR efficiency of >10 (W/kg)/W, nonuniformity of <40%, SAR variability of <17%, and a temperature rise of <0.13/spl deg/C/(W/kg) is achieved. The numerical dosimetry for the field and temperature distributions within the Petri dishes was verified using E-field and temperature probes. The temperature analysis has shown that the possibility of localized "hot spots" can be excluded.

[1]  L Laval,et al.  A new in vitro exposure device for the mobile frequency of 900 MHz. , 2000, Bioelectromagnetics.

[2]  国際非電離放射線防護委員会 ICNIRP statement on the "Guidelines for limiting exposure to time-varying electric, magnetic, and electromagnetic fields (up to 300 GHz)". , 2009, Health physics.

[3]  N Kuster,et al.  Design, optimization, realization, and analysis of an in vitro system for the exposure of embryonic stem cells at 1.71 GHz. , 2000, Bioelectromagnetics.

[4]  W. R. Adey,et al.  Exposure of nerve growth factor-treated PC12 rat pheochromocytoma cells to a modulated radiofrequency field at 836.55 MHz: effects on c-jun and c-fos expression. , 1997, Bioelectromagnetics.

[5]  Niels Kuster,et al.  Millimeter-resolution E-field probe for isotropic measurement in lossy media between 100 MHz and 20 GHz , 2000, IEEE Trans. Instrum. Meas..

[6]  N Kuster,et al.  Basis for optimization of in vitro exposure apparatus for health hazard evaluations of mobile communications , 2001, Bioelectromagnetics.

[7]  W F Pickard,et al.  The radial transmission line as a broad-band shielded exposure system for microwave irradiation of large numbers of culture flasks. , 1999, Bioelectromagnetics.

[8]  N. Kuster,et al.  In vitro exposure systems for RF exposures at 900 MHz , 2004, IEEE Transactions on Microwave Theory and Techniques.

[9]  P Regli,et al.  Electromagnetic and heat transfer computations for non-ionizing radiation dosimetry. , 2000, Physics in medicine and biology.

[10]  Dosimetry considerations in far field microwave exposure of mammalian cells. , 1988, Physiological chemistry and physics and medical NMR.

[11]  N. Kuster,et al.  Effect of the meniscus at the solid/liquid interface on the SAR distribution in Petri dishes and flasks , 2003, Bioelectromagnetics.

[12]  D Krause,et al.  Role of modulation on the effect of microwaves on ornithine decarboxylase activity in L929 cells. , 1997, Bioelectromagnetics.

[13]  Barry N. Taylor,et al.  Guidelines for Evaluating and Expressing the Uncertainty of Nist Measurement Results , 2017 .

[14]  J. Mcnamee,et al.  Cylindrical waveguide applicator for in vitro exposure of cell culture samples to 1.9‐GHz radiofrequency fields , 2002, Bioelectromagnetics.

[15]  N Kuster,et al.  Recommended minimal requirements and development guidelines for exposure setups of bio-experiments addressing the health risk concern of wireless communications. , 2000, Bioelectromagnetics.