Patient-specific radiation dose and cancer risk estimation in CT: part I. development and validation of a Monte Carlo program.

PURPOSE Radiation-dose awareness and optimization in CT can greatly benefit from a dose-reporting system that provides dose and risk estimates specific to each patient and each CT examination. As the first step toward patient-specific dose and risk estimation, this article aimed to develop a method for accurately assessing radiation dose from CT examinations. METHODS A Monte Carlo program was developed to model a CT system (LightSpeed VCT, GE Healthcare). The geometry of the system, the energy spectra of the x-ray source, the three-dimensional geometry of the bowtie filters, and the trajectories of source motions during axial and helical scans were explicitly modeled. To validate the accuracy of the program, a cylindrical phantom was built to enable dose measurements at seven different radial distances from its central axis. Simulated radial dose distributions in the cylindrical phantom were validated against ion chamber measurements for single axial scans at all combinations of tube potential and bowtie filter settings. The accuracy of the program was further validated using two anthropomorphic phantoms (a pediatric one-year-old phantom and an adult female phantom). Computer models of the two phantoms were created based on their CT data and were voxelized for input into the Monte Carlo program. Simulated dose at various organ locations was compared against measurements made with thermoluminescent dosimetry chips for both single axial and helical scans. RESULTS For the cylindrical phantom, simulations differed from measurements by -4.8% to 2.2%. For the two anthropomorphic phantoms, the discrepancies between simulations and measurements ranged between (-8.1%, 8.1%) and (-17.2%, 13.0%) for the single axial scans and the helical scans, respectively. CONCLUSIONS The authors developed an accurate Monte Carlo program for assessing radiation dose from CT examinations. When combined with computer models of actual patients, the program can provide accurate dose estimates for specific patients.

[1]  William E. Lorensen,et al.  Marching cubes: a high resolution 3D surface construction algorithm , 1996 .

[2]  William H. Judy,et al.  The Measurement, Reporting, and Management of Radiation Dose in CT , 2008 .

[3]  D R Dance,et al.  CT dosimetry: getting the best from the adult Cristy phantom. , 2005, Radiation protection dosimetry.

[4]  Willi A. Kalender,et al.  Validation of a Monte Carlo tool for patient-specific dose simulations in multi-slice computed tomography , 2008, European Radiology.

[5]  W P Segars,et al.  Realistic CT simulation using the 4D XCAT phantom. , 2008, Medical physics.

[6]  J. Sempau,et al.  PENELOPE-2006: A Code System for Monte Carlo Simulation of Electron and Photon Transport , 2009 .

[7]  J J DeMarco,et al.  A Monte Carlo-based method to estimate radiation dose from spiral CT: from phantom testing to patient-specific models , 2003, Physics in medicine and biology.

[8]  Nina Petoussi-Henss,et al.  The GSF family of voxel phantoms , 2002, Physics in medicine and biology.

[9]  D. Broga,et al.  Ionizing Radiation Exposure of the Population of the United States , 2009 .

[10]  Wesley E Bolch,et al.  Organ and effective doses in pediatric patients undergoing helical multislice computed tomography examination. , 2007, Medical physics.

[11]  P Andreo,et al.  Configuration of the electron transport algorithm of PENELOPE to simulate ion chambers , 2006, Physics in medicine and biology.

[12]  Cynthia H McCollough,et al.  The feasibility of a scanner-independent technique to estimate organ dose from MDCT scans: using CTDIvol to account for differences between scanners. , 2010, Medical physics.

[13]  Jacob Geleijns,et al.  Overranging in multisection CT: quantification and relative contribution to dose--comparison of four 16-section CT scanners. , 2007, Radiology.

[14]  J J DeMarco,et al.  Estimating radiation doses from multidetector CT using Monte Carlo simulations: effects of different size voxelized patient models on magnitudes of organ and effective dose , 2007, Physics in medicine and biology.

[15]  T. S. Chen,et al.  Patient risk from interproximal radiography. , 1984, Oral surgery, oral medicine, and oral pathology.

[16]  W. Paul Segars,et al.  Patient-specific radiation dose and cancer risk estimation in CT: part II. Application to patients. , 2010, Medical physics.

[17]  Wesley E Bolch,et al.  Organ and effective doses in newborn patients during helical multislice computed tomography examination , 2006, Physics in medicine and biology.

[18]  M Caon,et al.  An EGS4-ready tomographic computational model of a 14-year-old female torso for calculating organ doses from CT examinations. , 1999, Physics in medicine and biology.

[19]  P C Shrimpton,et al.  Influence of patient age on normalized effective doses calculated for CT examinations. , 2002, The British journal of radiology.

[20]  Donald P. Frush,et al.  The ‘Image Gently’ campaign: increasing CT radiation dose awareness through a national education and awareness program , 2008, Pediatric Radiology.

[21]  Maria Zankl,et al.  Organ Doses for Children from Computed Tomographic Examinations , 1995 .

[22]  Michael J Pentecost,et al.  American College of Radiology white paper on radiation dose in medicine. , 2007, Journal of the American College of Radiology : JACR.

[23]  J. Sempau,et al.  Experimental benchmarks of the Monte Carlo code penelope , 2003 .

[24]  J. Baró,et al.  PENELOPE: An algorithm for Monte Carlo simulation of the penetration and energy loss of electrons and positrons in matter , 1995 .

[25]  William Schroeder,et al.  The Visualization Toolkit: An Object-Oriented Approach to 3-D Graphics , 1997 .

[26]  Wesley E Bolch,et al.  Whole-body voxel phantoms of paediatric patients—UF Series B , 2006, Physics in medicine and biology.

[27]  W Abmayr,et al.  The calculations of dose from external photon exposures using reference and realistic human phantoms and Monte Carlo methods , 1986 .

[28]  Maria Zankl,et al.  Reducing radiation dose to selected organs by selecting the tube start angle in MDCT helical scans: a Monte Carlo based study. , 2009, Medical physics.

[29]  J J DeMarco,et al.  A Monte Carlo based method to estimate radiation dose from multidetector CT (MDCT): cylindrical and anthropomorphic phantoms. , 2005, Physics in medicine and biology.

[30]  Daniel Lodwick,et al.  Hybrid computational phantoms of the 15-year male and female adolescent: applications to CT organ dosimetry for patients of variable morphometry. , 2008, Medical physics.