ACCURACY OF ELECTROMAGNETIC TRACKING IN AN IMAGE-GUIDED SURGERY SUITE

Recently, electromagnetic tracking for surgical procedures has gained popularity due to its small sensor size and the absence of line-of-sight restrictions. However, EM trackers are susceptible to measurement noise. Indeed, depending on the environment, measurement uncertainties may vary considerably. Therefore, it is important to characterise electromagnetic measurement systems when used in a fluoroscopy setting. The purpose of our study is to assess decoupled static electromagnetic measurement errors in position and orientation, without adding potential interference, in the presence of fluoroscopic imaging equipment.Using an Aurora electromagnetic tracking system (Northern Digital, Waterloo, Canada), 5 degrees of freedom measurements were collected in a working space located midway between the source and the receiver of a flat-panel 3D fluoroscope (Innova 4100, GE Healthcare, Buc, France) emitting X-rays. In addition, to determine potential EM distortion from X-rays, electromagnetic measurement accuraci...