Mechanical testing of internal fixation devices: A theoretical and practical examination of current methods.

Successful healing of long bone fractures is dependent on the mechanical environment created within the fracture, which in turn is dependent on the fixation strategy. Recent literature reports have suggested that locked plating devices are too stiff to reliably promote healing. However, in vitro testing of these devices has been inconsistent in both method of constraint and reported outcomes, making comparisons between studies and the assessment of construct stiffness problematic. Each of the methods previously used in the literature were assessed for their effect on the bending of the sample and concordant stiffness. The choice of outcome measures used in in vitro fracture studies was also assessed. Mechanical testing was conducted on seven hole locked plated constructs in each method for comparison. Based on the assessment of each method the use of spherical bearings, ball joints or similar is suggested at both ends of the sample. The use of near and far cortex movement was found to be more comprehensive and more accurate than traditional centrally calculated interfragmentary movement values; stiffness was found to be highly susceptible to the accuracy of deformation measurements and constraint method, and should only be used as a within study comparison method. The reported stiffness values of locked plate constructs from in vitro mechanical testing is highly susceptible to testing constraints and output measures, with many standard techniques overestimating the stiffness of the construct. This raises the need for further investigation into the actual mechanical behaviour within the fracture gap of these devices.

[1]  R. Nanda,et al.  Biomechanical testing of the locking compression plate: when does the distance between bone and implant significantly reduce construct stability? , 2007, Injury.

[2]  L. Claes,et al.  Influence of size and stability of the osteotomy gap on the success of fracture healing , 1997, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[3]  V Vécsei,et al.  A new modular testing system for biomechanical evaluation of tibial intramedullary fixation devices. , 2001, Injury.

[4]  Georg N. Duda,et al.  Bewertung der Stabilität von Frakturfixationssystemen: Mechanische Vorrichtung zur Untersuchung der 3-D-Steifigkeit in vitro - Assessment of the Stability of Fracture Fixation Systems: Mechanical Device to Investigate the 3-D Stiffness in vitro , 2001 .

[5]  Michael Bottlang,et al.  Relative stability of conventional and locked plating fixation in a model of the osteoporotic femoral diaphysis. , 2009, Clinical biomechanics.

[6]  Georg N Duda,et al.  Mechanical conditions in the internal stabilization of proximal tibial defects. , 2002, Clinical biomechanics.

[7]  M. Bottlang,et al.  Biomechanics of Far Cortical Locking , 2011, Journal of orthopaedic trauma.

[8]  Ulrich Dieter,et al.  Biomechanical testing of the LCP--how can stability in locked internal fixators be controlled? , 2003, Injury.

[9]  P. Augat,et al.  Interfragmentary Movement in Diaphyseal Tibia Fractures Fixed With Locked Intramedullary Nails , 2008, Journal of orthopaedic trauma.

[10]  Georg N Duda,et al.  Timely fracture-healing requires optimization of axial fixation stability. , 2007, The Journal of bone and joint surgery. American volume.

[11]  M. Gardner,et al.  Stiffness Modulation of Locking Plate Constructs Using Near Cortical Slotted Holes: A Preliminary Study , 2009, Journal of orthopaedic trauma.

[12]  P. Augat,et al.  Effects of construct stiffness on healing of fractures stabilized with locking plates. , 2010, The Journal of bone and joint surgery. American volume.

[13]  Martyn Snow,et al.  A Mechanical Comparison of the Locking Compression Plate (LCP) and the Low Contact-Dynamic Compression Plate (DCP) in an Osteoporotic Bone Model , 2008, Journal of orthopaedic trauma.

[14]  L Claes,et al.  A method to determine the 3-D stiffness of fracture fixation devices and its application to predict inter-fragmentary movement. , 1997, Journal of biomechanics.

[15]  M. Gardner,et al.  Less rigid stable fracture fixation in osteoporotic bone using locked plates with near cortical slots. , 2010, Injury.

[16]  S. Stover,et al.  Mechanical comparison of 3.5 mm broad dynamic compression plate, broad limited-contact dynamic compression plate, and narrow locking compression plate systems using interfragmentary gap models. , 2008, Veterinary surgery : VS.

[17]  Klaus Püschel,et al.  Influence of intramedullary nail diameter and locking mode on the stability of tibial shaft fracture fixation , 2009, Archives of Orthopaedic and Trauma Surgery.

[18]  Antonio Meleddu,et al.  A rigorous method for evaluation of the 6D compliance of external fixators , 2007, Biomechanics and modeling in mechanobiology.

[19]  W. F. Riley,et al.  Statics and Mechanics of Materials: An Integrated Approach , 1996 .

[20]  M. Bottlang,et al.  A nonlocking end screw can decrease fracture risk caused by locked plating in the osteoporotic diaphysis. , 2009, The Journal of bone and joint surgery. American volume.