A Microfluidic Device for Simultaneous Extraction of Plasma, Red Blood Cells, and On-Chip White Blood Cell Trapping

This study reports a microfluidic device for whole blood processing. The device uses the bifurcation law, cross-flow method, and hydrodynamic flow for simultaneous extraction of plasma, red blood cells, and on-chip white blood cell trapping. The results demonstrate successful plasma and red blood cell collection with a minimum dilution factor (0.76x) and low haemolysis effect. The extracted red blood cells can also be applied for blood type tests. Moreover, the device can trap up to ~1,800 white blood cells in 20 minutes. The three components can be collected simultaneously using only 6 μL of whole blood without any sample preparation processes. Based on these features, the microfluidic device enables low-cost, rapid, and efficient whole blood processing functionality that could potentially be applied for blood analysis in resource-limited environments or point-of-care settings.

[1]  A. B. Frazier,et al.  Paramagnetic capture mode magnetophoretic microseparator for high efficiency blood cell separations. , 2006, Lab on a chip.

[2]  E. Chuang,et al.  An automated microfluidic DNA microarray platform for genetic variant detection in inherited arrhythmic diseases. , 2018, The Analyst.

[3]  Y. Fung,et al.  Effect of velocity of distribution on red cell distribution in capillary blood vessels. , 1978, The American journal of physiology.

[4]  E. Sollier,et al.  Fast and continuous plasma extraction from whole human blood based on expanding cell-free layer devices , 2010, Biomedical microdevices.

[5]  Yen-Hung Lin,et al.  A microfluidic device integrating dual CMOS polysilicon nanowire sensors for on-chip whole blood processing and simultaneous detection of multiple analytes. , 2016, Lab on a chip.

[6]  Sung Yang,et al.  A microfluidic device for continuous white blood cell separation and lysis from whole blood. , 2010, Artificial organs.

[7]  T. Laurell,et al.  Review of cell and particle trapping in microfluidic systems. , 2009, Analytica chimica acta.

[8]  Han Wei Hou,et al.  Microfluidic Devices for Blood Fractionation , 2011, Micromachines.

[9]  F. Marley Taking the Blood Apart , 1968 .

[10]  H. Ji,et al.  Silicon-based microfilters for whole blood cell separation , 2008, Biomedical microdevices.

[11]  K. Kotz,et al.  Acoustic separation in plastic microfluidics for rapid detection of bacteria in blood using engineered bacteriophage. , 2018, Lab on a chip.

[12]  Yong-Jun Kim,et al.  Hybrid cell sorters for on-chip cell separation by hydrodynamics and magnetophoresis , 2010 .

[13]  Luke P. Lee,et al.  Stand-alone self-powered integrated microfluidic blood analysis system (SIMBAS). , 2011, Lab on a chip.

[14]  M. Harboe,et al.  A method for determination of hemoglobin in plasma by near-ultraviolet spectrophotometry. , 1959, Scandinavian journal of clinical and laboratory investigation.

[15]  Dennie V. Jones,et al.  The value of a complete blood count in predicting cancer of the colon. , 2004, Cancer detection and prevention.

[16]  Arnan Mitchell,et al.  Microfluidic platform for separation and extraction of plasma from whole blood using dielectrophoresis. , 2015, Biomicrofluidics.

[17]  Jasmina Casals-Terré,et al.  Hydrodynamic and direct-current insulator-based dielectrophoresis (H-DC-iDEP) microfluidic blood plasma separation , 2015, Analytical and Bioanalytical Chemistry.

[18]  Ion Stiharu,et al.  Interdigitated comb‐like electrodes for continuous separation of malignant cells from blood using dielectrophoresis , 2011, Electrophoresis.

[19]  Mengxi Wu,et al.  High-throughput acoustic separation of platelets from whole blood. , 2016, Lab on a chip.

[20]  J. Ducrée,et al.  Rapid, low-cost and instrument-free CD4+ cell counting for HIV diagnostics in resource-poor settings. , 2014, Lab on a chip.

[21]  Yong Li,et al.  Microfluidic disk for the determination of human blood types , 2017 .

[22]  R. Christensen,et al.  The CBC: reference ranges for neonates. , 2009, Seminars in perinatology.

[23]  E. Sollier,et al.  Micro-scale blood plasma separation: from acoustophoresis to egg-beaters. , 2013, Lab on a chip.

[24]  Luke P. Lee,et al.  Self-powered integrated microfluidic point-of-care low-cost enabling (SIMPLE) chip , 2017, Science Advances.

[25]  Dino Di Carlo,et al.  Microfluidic sample preparation for diagnostic cytopathology. , 2013, Lab on a chip.

[26]  J. Thomson,et al.  Dielectrophoretic separation of platelets from diluted whole blood in microfluidic channels , 2008, Electrophoresis.

[27]  A. Undar,et al.  A microfluidic device for continuous, real time blood plasma separation. , 2006, Lab on a chip.

[28]  S. Tripathi,et al.  Microdevice for plasma separation from whole human blood using bio-physical and geometrical effects , 2016, Scientific Reports.

[29]  Mary Louise Turgeon,et al.  Clinical Hematology: Theory and Procedures , 1988 .

[30]  J. Sturm,et al.  Deterministic hydrodynamics: Taking blood apart , 2006, Proceedings of the National Academy of Sciences.

[31]  Eugene J. Lim,et al.  Microfluidic, marker-free isolation of circulating tumor cells from blood samples , 2014, Nature Protocols.

[32]  Chien-Fu Chen,et al.  Rapid and inexpensive blood typing on thermoplastic chips. , 2015, Lab on a chip.

[33]  Hansen Bow,et al.  Microfluidics for cell separation , 2010, Medical & Biological Engineering & Computing.

[34]  R. Skalak,et al.  Cell distribution in capillary networks. , 1980, Microvascular research.

[35]  Benjamin B Yellen,et al.  Magnetic separation of acoustically focused cancer cells from blood for magnetographic templating and analysis. , 2016, Lab on a chip.

[36]  Thomas Laurell,et al.  Acoustofluidic, label-free separation and simultaneous concentration of rare tumor cells from white blood cells. , 2015, Analytical chemistry.

[37]  Y. Fung Stochastic flow in capillary blood vessels. , 1973, Microvascular research.