Frequency response calibration of recess-mounted pressure transducers

A technique is described for measuring the frequency response of pressure transducers mounted inside a model, where a narrow pipette leads to an orifice at the surface. An acoustic driver is mounted to a small chamber which has an opening at the opposite end with an O-ring seal to place over the orifice. A 3.18 mm (1/8 inch) reference microphone is mounted to one side of the chamber. The acoustic driver receives an input of white noise, and the transducer and reference microphone outputs are compared to obtain the frequency response of the pressure transducer. Selected results are presented in the form of power spectra for both the transducer and the reference, as well as the amplitude variation and phase shift between the two signals as a function of frequency. The effect of pipette length and the use of this technique for identifying both blocked orifices and faulty transducers are described.