Performance evaluation of a subset of a four-layer LSO detector for a small animal DOI PET scanner: jPET-RD

Previously, we proposed a new depth of interaction (DOI) encoding method and proved that it worked successfully with four-layered Gd/sub 2/SiO/sub 5/ crystals for a small animal positron emission tomography (PET) detector. We are now planning to develop a small animal PET scanner, jPET-RD (for rodents with DOI detectors), which has both high resolution and high sensitivity by the use of a DOI detector with a 32/spl times/32/spl times/4 crystal array. The scintillator for the detector will be Lu/sub 2(1-x)/Y/sub 2x/SiO/sub 5/ (LYSO). In this work, we evaluated performance of a DOI detector composed of four layers of a 12/spl times/12 LYSO (Lu: 98%, Y: 2%) crystal array by irradiating 511 keV gamma rays uniformly. The new encoding method was used for crystal identification. The size of each crystal was 1.46 mm/spl times/1.46 mm/spl times/4.5 mm. The crystal block was coupled to a 256-channel flat panel position sensitive photomultiplier tube, which has 16/spl times/16 multi anodes at intervals of 3.04 mm. As we expected, all crystals are expressed on a single two-dimensional position histogram without overlapping. Energy resolution of all events is 21.8% and time resolution of all events is 0.69 ns in FWHM. When layers are counted from the top, the energy resolutions of the first, second, third, and fourth layer events are 11.6%, 12.3%, 13.3%, and 19.1% and the time resolutions are 0.60ns, 0.59ns, 0.60ns, and 0.66ns, respectively.

[1]  H. Murayama,et al.  Performance of 256ch flat panel PS-PMT with small crystals for a DOI PET detector , 2003, 2003 IEEE Nuclear Science Symposium. Conference Record (IEEE Cat. No.03CH37515).

[2]  H. Kawai,et al.  Performance of a PET detector with a 256ch flat panel PS-PMT , 2004, IEEE Transactions on Nuclear Science.

[3]  H. Murayama,et al.  BGO DOI-detector composed of three-dimensional crystal array for PET , 2004, IEEE Symposium Conference Record Nuclear Science 2004..

[4]  H. Murayama,et al.  A four-layer depth of interaction detector block for small animal PET , 2003, 2003 IEEE Nuclear Science Symposium. Conference Record (IEEE Cat. No.03CH37515).

[5]  H. Murayama,et al.  Preliminary design studies of a high sensitivity small animal DOI-PET scanner: jPET-RD , 2004, IEEE Symposium Conference Record Nuclear Science 2004..

[6]  H. Kawai,et al.  Performance of a PET detector with a 256ch flat panel PS-PMT , 2002 .