Analysis of the artifact reduction rate for the types of medical metals in CT with MAR algorithm

We investigated on the usefulness of MAR algorithm by making a comparison of the CT value between before and after applying the MAR algorithm in dual energy CT, using the various kinds of medical metals, causing the artifact to lead to the low image quality. As a result, the artifact was reduced in most cases (P<0.05); in particular, the artifact was highly reduced (P<0.05) using high density material, like alloy-stainless (reduced by 78.1%) and platinum, for example GDC coil (reduced by 76.1%). The effect of decreasing the Black hole artifact was outstanding in both the alloy-stainless and alloy-titanium (P<0.05). However, in case of GDC coil-a type platinum, white streak artifact was reduced effectively (P<0.05). Therefore, in case of patients who have medical metals inserted, we think that high-quality image information can be provided by decreasing the artifact caused by high density material through MAR algorithm in dual energy CT.