Implantation of the VentrAssist Implantable Rotary Blood Pump in Sheep

&NA; The VentrAssist Implantable Rotary Blood Pump (IRBP) is a hydrodynamically suspended, electromagnetically driven, centrifugal blood pump that provides continuous flow of up to 10 L/min at 3,000 rpm. In vivo studies in sheep were conducted to assess system design and performance. Surgery involved thoracotomy with subdiaphragmatic pump placement. Cannulae were transdiaphragmatic, with inflow in the left ventricular apex and outflow anastomosed to the descending aorta. Animals had no anticoagulation or antiplatelet therapy after surgery and no prophylactic antibiotics after recovery. Twelve sheep were supported for 622 pump days. Estimated pump flow ranged from 1 to 5.5 L/min at 1,800 to 2,000 rpm using 2.5 to 4.5 W. There was no clinical evidence of hemolysis or cardiovascular, renal, or hepatic dysfunction. Adverse outcomes included kinking/disconnection of the outflow cannula caused by the graft bend relief (n 4), which was addressed through cannula redesign. Pump electrical malfunction (n 4), caused by a silicone potting compound, was corrected using a neutral curing potting material. Surgical/husbandry issues (n 2) also were addressed. The VentrAssist IRBP provides high flow at low rotational speed and power consumption. Further trials are in progress in advance of in vivo studies of the safety and efficacy of the final system. ASAIO Journal 2003; 49:454‐458.