Comparative push-out test of dense HA implants and HA-coated implants: findings in a canine study.

Two types of hydroxyapatite (HA) implants have been developed: an HA-coated implant and a dense HA implant. For a longer in situ life span, the HA implant must remain chemically stable and possess high resistance to occlusal force. To determine which type of HA implant shows better durability, this comparative dog study was done to evaluate push-out test results of HA-coated implants and dense HA implants of approximately the same size after implantation in the mandibular and coxal bones for periods ranging from 3 weeks to 10 months. The findings revealed that for the mandibular implants, the push-out values of HA-coated implants were significantly higher than those of dense HA implants at 2 and 4 months after implantation, with significance levels of p < .001 and p < 0.05, respectively. However, there was no significant difference between the two implant types at 10 months. As for the coxal implants, no significant differences were noted for any period. Furthermore, the ratio of push-out values of the dense HA implants to those of the HA-coated implants situated in the same position bilaterally in each bone of the body for each implantation period rose with the passage of time, especially in the mandible. In the mandibular implants, the correlation coefficient of the relationship between the ratio and duration of implantation was highly significant (p < 0.001). Push-out testing caused detachment of the surface portion of the HA coating that was bound to the dense bone from the HA-coated implant at 2, 4, and 10 months after implantation. Furthermore, at 10 months the HA-coated layer in the wide areas of the implants had completely detached from the metal substrate, in contrast to the dense HA implants, which remained durable throughout the test period.

[1]  Y. Yamashita,et al.  Microstructural changes in bone of HA-coated implants. , 1998, Journal of biomedical materials research.

[2]  N. Nakabayashi,et al.  Adhesive improvement of the mechanical properties of a dense HA-cemented Ti dental implant. , 1996, Journal of biomedical materials research.

[3]  C. Klein,et al.  Calcium phosphate plasma-sprayed coatings and their stability: an in vivo study. , 1994, Journal of biomedical materials research.

[4]  E. van der Velde,et al.  In vivo tensile testing of fluorapatite and hydroxylapatite plasma-sprayed coatings. , 1994, Journal of biomedical materials research.

[5]  M. Ogiso,et al.  Clinical evaluation of two-piece apatite dental implants in 189 cases. , 1993, Journal of long-term effects of medical implants.

[6]  S. Radin,et al.  The effect of calcium phosphate ceramic composition and structure on in vitro behavior. I. Dissolution. , 1993, Journal of biomedical materials research.

[7]  C. B. Ponton,et al.  Microstructural characterization of hydroxyapatite coating on titanium , 1992 .

[8]  Cook Sd,et al.  Relationship between surface characteristics and the degree of bone-implant integration. , 1992, Journal of biomedical materials research.

[9]  P. Ducheyne,et al.  Plasma spraying induced changes of calcium phosphate ceramic characteristics and the effect onin vitro stability , 1992 .

[10]  T. Ichijo,et al.  Examination of human bone surrounded by a dense hydroxyapatite dental implant after long-term use. , 1992, Journal of long-term effects of medical implants.

[11]  P. Rozing,et al.  A finite element analysis of the push-out test: influence of test conditions. , 1992, Journal of biomedical materials research.

[12]  M. Jarcho Retrospective analysis of hydroxyapatite development for oral implant applications. , 1992, Dental clinics of North America.

[13]  K. de Groot,et al.  X-ray diffraction studies on plasma-sprayed calcium phosphate-coated implants. , 1990, Journal of biomedical materials research.

[14]  M. Block,et al.  The effect of diameter and length of hydroxylapatite-coated dental implants on ultimate pullout force in dog alveolar bone. , 1990, Journal of oral and maxillofacial surgery : official journal of the American Association of Oral and Maxillofacial Surgeons.

[15]  J Black,et al.  "Push-out" tests. , 1989, Journal of biomedical materials research.

[16]  C. Klein,et al.  Plasma sprayed coatings of hydroxylapatite. , 1987, Journal of biomedical materials research.

[17]  S D Cook,et al.  Interface mechanics and histology of titanium and hydroxylapatite-coated titanium for dental implant applications. , 1987, The International journal of oral & maxillofacial implants.

[18]  R. Doremus,et al.  Direct electron microscopy studies of the bone-hydroxylapatite interface. , 1984, Journal of biomedical materials research.