Systematic calibration of an integrated x-ray and optical tomography system for preclinical radiation research.

PURPOSE The cone beam computed tomography (CBCT) guided small animal radiation research platform (SARRP) has been developed for focal tumor irradiation, allowing laboratory researchers to test basic biological hypotheses that can modify radiotherapy outcomes in ways that were not feasible previously. CBCT provides excellent bone to soft tissue contrast, but is incapable of differentiating tumors from surrounding soft tissue. Bioluminescence tomography (BLT), in contrast, allows direct visualization of even subpalpable tumors and quantitative evaluation of tumor response. Integration of BLT with CBCT offers complementary image information, with CBCT delineating anatomic structures and BLT differentiating luminescent tumors. This study is to develop a systematic method to calibrate an integrated CBCT and BLT imaging system which can be adopted onboard the SARRP to guide focal tumor irradiation. METHODS The integrated imaging system consists of CBCT, diffuse optical tomography (DOT), and BLT. The anatomy acquired from CBCT and optical properties acquired from DOT serve as a priori information for the subsequent BLT reconstruction. Phantoms were designed and procedures were developed to calibrate the CBCT, DOT/BLT, and the entire integrated system. Geometrical calibration was performed to calibrate the CBCT system. Flat field correction was performed to correct the nonuniform response of the optical imaging system. Absolute emittance calibration was performed to convert the camera readout to the emittance at the phantom or animal surface, which enabled the direct reconstruction of the bioluminescence source strength. Phantom and mouse imaging were performed to validate the calibration. RESULTS All calibration procedures were successfully performed. Both CBCT of a thin wire and a euthanized mouse revealed no spatial artifact, validating the accuracy of the CBCT calibration. The absolute emittance calibration was validated with a 650 nm laser source, resulting in a 3.0% difference between simulated and measured signal. The calibration of the entire system was confirmed through the CBCT and BLT reconstruction of a bioluminescence source placed inside a tissue-simulating optical phantom. Using a spatial region constraint, the source position was reconstructed with less than 1 mm error and the source strength reconstructed with less than 24% error. CONCLUSIONS A practical and systematic method has been developed to calibrate an integrated x-ray and optical tomography imaging system, including the respective CBCT and optical tomography system calibration and the geometrical calibration of the entire system. The method can be modified and adopted to calibrate CBCT and optical tomography systems that are operated independently or hybrid x-ray and optical tomography imaging systems.

[1]  Hamid Dehghani,et al.  Fast segmentation and high-quality three-dimensional volume mesh creation from medical images for diffuse optical tomography , 2013, Journal of biomedical optics.

[2]  J H Siewerdsen,et al.  Geometric calibration of a mobile C-arm for intraoperative cone-beam CT. , 2008, Medical physics.

[3]  Ralph Weissleder,et al.  Molecular optical imaging: Applications leading to the development of present day therapeutics , 2005, NeuroRX.

[4]  Michael S. Patterson,et al.  Improved bioluminescence and fluorescence reconstruction algorithms using diffuse optical tomography, normalized data, and optimized selection of the permissible source region , 2010, Biomedical optics express.

[5]  V. Ntziachristos Fluorescence molecular imaging. , 2006, Annual review of biomedical engineering.

[6]  R Clackdoyle,et al.  Analytic method based on identification of ellipse parameters for scanner calibration in cone-beam tomography. , 2000, Physics in medicine and biology.

[7]  Ping Wu,et al.  Bioluminescence Tomography by an Iterative Reweighted ${\bm {l_{2}}}$-Norm Optimization , 2014, IEEE Transactions on Biomedical Engineering.

[8]  Drew M. Pardoll,et al.  Adjuvant Vaccine Therapy in Combination with Radiation for Treatment of Prostate Cancer: Preclinical Validation in a Mouse Model of Prostate Cancer , 2008 .

[9]  Ali Behrooz,et al.  Adaptive row-action inverse solver for fast noise-robust three-dimensional reconstructions in bioluminescence tomography: theory and dual-modality optical/computed tomography in vivo studies , 2013, Journal of biomedical optics.

[10]  M. Patterson,et al.  Algorithms for bioluminescence tomography incorporating anatomical information and reconstruction of tissue optical properties , 2010, Biomedical optics express.

[11]  B. Pogue,et al.  Spectrally resolved bioluminescence optical tomography. , 2006, Optics letters.

[12]  William M Gallagher,et al.  Bioluminescent imaging: a critical tool in pre‐clinical oncology research , 2010, The Journal of pathology.

[13]  Michael S. Patterson,et al.  Fabrication and characterization of phantoms with tissue-like optical properties from 500 to 700 nm , 2010 .

[14]  Ping Wu,et al.  Bioluminescence tomography by an iterative reweighted (l)2 norm optimization. , 2014, IEEE transactions on bio-medical engineering.

[15]  D. Jenkins,et al.  Bioluminescent human breast cancer cell lines that permit rapid and sensitive in vivo detection of mammary tumors and multiple metastases in immune deficient mice , 2005, Breast Cancer Research.

[16]  Kai Yang,et al.  A geometric calibration method for cone beam CT systems. , 2006, Medical physics.

[17]  Hamid Dehghani,et al.  In vivo bioluminescence tomography with a blocking-off finite-difference SP3 method and MRI/CT coregistration. , 2009, Medical physics.

[18]  C. Alberti,et al.  From molecular imaging in preclinical/clinical oncology to theranostic applications in targeted tumor therapy. , 2012, European review for medical and pharmacological sciences.

[19]  Ge Wang,et al.  Overview of bioluminescence tomography--a new molecular imaging modality. , 2008, Frontiers in bioscience : a journal and virtual library.

[20]  Erik Tryggestad,et al.  Small animal radiotherapy research platforms , 2011, Physics in medicine and biology.

[21]  I Iordachita,et al.  An integrated x-ray/optical tomography system for pre-clinical radiation research , 2013, Medical Imaging.

[22]  Michael S. Patterson,et al.  Self-calibrated algorithms for diffuse optical tomography and bioluminescence tomography using relative transmission images , 2012, Biomedical optics express.

[23]  Hongjun Song,et al.  Localized CT-Guided Irradiation Inhibits Neurogenesis in Specific Regions of the Adult Mouse Brain , 2011, Radiation research.

[24]  Young-Bin Cho,et al.  Geometric Calibration of Novel Cone-Beam CT Developed For Dedicated Stereotactic Radiotherapy Unit , 2011 .

[25]  F Verhaegen,et al.  A comprehensive system for dosimetric commissioning and Monte Carlo validation for the small animal radiation research platform , 2009, Physics in medicine and biology.

[26]  Hamid Dehghani,et al.  Near infrared optical tomography using NIRFAST: Algorithm for numerical model and image reconstruction. , 2009, Communications in numerical methods in engineering.

[27]  E. Hoffman,et al.  In vivo mouse studies with bioluminescence tomography. , 2006, Optics express.

[28]  Peter Kazanzides,et al.  High-resolution, small animal radiation research platform with x-ray tomographic guidance capabilities. , 2008, International journal of radiation oncology, biology, physics.

[29]  Vasilis Ntziachristos,et al.  Looking and listening to light: the evolution of whole-body photonic imaging , 2005, Nature Biotechnology.

[30]  Erik Tryggestad,et al.  Differentiation between glioma and radiation necrosis using molecular magnetic resonance imaging of endogenous proteins and peptides , 2010, Nature Medicine.

[31]  Jie Tian,et al.  Incorporating MRI structural information into bioluminescence tomography: system, heterogeneous reconstruction and in vivo quantification. , 2014, Biomedical optics express.

[32]  Domingos Vieira,et al.  Preclinical Imaging: an Essential Ally in Modern Biosciences , 2013, Molecular Diagnosis & Therapy.

[33]  E. Galun,et al.  Imaging transgene expression in live animals. , 2001, Molecular therapy : the journal of the American Society of Gene Therapy.

[34]  Scott C Davis,et al.  Pre-clinical whole-body fluorescence imaging: Review of instruments, methods and applications. , 2010, Journal of photochemistry and photobiology. B, Biology.

[35]  John Wong,et al.  Microdialysis to Measure Radiation-induced Blood-brain Barrier Disruption in Mice , 2008 .