Dual-energy, standard and low-kVp contrast-enhanced CT-cholangiography: a comparative analysis of image quality and radiation exposure.

OBJECTIVE Quantitative image quality assessment in terms of image noise (IN), contrast-to-noise ratio (CNR), and signal-to-noise ratio (SNR) in relation to required radiation dose (RD) for dual-energy (DE), standard and low-kVp contrast-enhanced computed-tomography (CT) cholangiography. MATERIALS AND METHODS For each of 22 DECT-cholangiography examinations, 3 image datasets were analyzed as independent single-source CT-acquisitions at different tube potential, i.e. 80 kVp, 120 kVp-equivalent (linear blended dataset M0.3: 30% 80 kVp, 70% 140 kVp), and 140 kVp. Analysis comprised determination of IN, CNR and SNR in regions of interest (ROI) placed in liver parenchyma and contrasted bile ducts. IN was evaluated as mean standard deviation of 3 ROI placed within liver parenchyma (segments 6/7, 5/8, 2/3); CNR was assessed as bile duct-to-liver parenchyma ratio, and SNR as bile duct-to-image noise ratio. RD in terms of CT dose index (CTDI(vol)), dose-length product (DLP) and effective dose (ED) has been determined for each of the datasets, and compared to console prediction and scan summary values. Using phantom measurements of CTDI(vol), a method for separating comprehensive RD values of DE-acquisitions into the original RD contribution of each tube (80 kVp/140 kVp) has been developed, enabling comparison of all 3 datasets as if independently acquired using single-source "single-energy" technique. RESULTS Highest IN was detected for 80 kVp- (38.6 ± 5.1HU), lowest for 120 kVp-equivalent linear blended M0.3-datasets (23.1 ± 3.4HU) with significant differences between all datasets (P<0.001). Highest SNR and CNR were measured for M0.3- (SNR: 14.8 ± 4.1; CNR: 11.6 ± 3.8) and 80 kVp-datasets (SNR: 13.8 ± 4.8; CNR: 11.2 ± 4.5); lowest for 140 kVp-datasets (SNR: 9.5 ± 2.5; CNR: 7.1 ± 2.3) with significant differences between M0.3- and 140 kVp-datasets as well as between 80 kVp- and 140 kVp-datasets (both P<0.001 for both CNR, SNR). CTDI(vol), DLP and ED were reduced by 50% for low-kilovoltage acquisitions (CTDI(vol): 5.5 ± 1.4 mGy; DLP: 127.8 ± 40.1 mGy cm; ED: 1.9 ± 0.6 mSv) compared to comprehensive DE-acquisitions (CTDI(vol): 11.0 ± 2.3 mGy; DLP: 253.8 ± 67.5 mGy cm; ED: 3.8 ± 1.0 mSv, tube contribution: 80 kVp: 44.5%; 140 kVp: 55.5%), and by 20% compared to conventional acquisitions at 120 kVp (CTDI(vol): 6.71 mGy; DLP: 153.5 ± 16.9 mGy cm; ED: 2.3 ± 0.3 mSv). CONCLUSIONS Despite higher IN, low-kilovoltage CT-cholangiography reveals no significant difference with respect to CNR and SNR when compared to linear blended images yielded by DECT. Compared to DECT or conventional CT at 120 kVp, contrast-enhanced low-kVp CT cholangiography potentially allows reduction of patient dose by up to 50% or 20%, respectively. Therefore, CT-cholangiography at 80 kVp should be considered as an alternative to DECT-cholangiography whenever DECT is unavailable, or if increased image quality of DECT regarding quantitative bile duct evaluation is not needed for diagnosis.

[1]  K. Stierstorfer,et al.  Technical principles of dual source CT. , 2008, European journal of radiology.

[2]  B. Yeh,et al.  Dual-energy and low-kVp CT in the abdomen. , 2009, AJR. American journal of roentgenology.

[3]  Takeshi Nakaura,et al.  Abdominal CT with low tube voltage: preliminary observations about radiation dose, contrast enhancement, image quality, and noise. , 2005, Radiology.

[4]  Jae Young Lee,et al.  Added value of 80 kVp images to averaged 120 kVp images in the detection of hepatocellular carcinomas in liver transplantation candidates using dual-source dual-energy MDCT: results of JAFROC analysis. , 2011, European journal of radiology.

[5]  Lee W Goldman,et al.  Principles of CT: Radiation Dose and Image Quality* , 2007, Journal of Nuclear Medicine Technology.

[6]  K. Stierstorfer,et al.  First performance evaluation of a dual-source CT (DSCT) system , 2006, European Radiology.

[7]  Peter Vock,et al.  Patient Exposure and Image Quality of Low-Dose Pulmonary Computed Tomography Angiography: Comparison of 100- and 80-kVp Protocols , 2008, Investigative radiology.

[8]  W Huda,et al.  Technique factors and image quality as functions of patient weight at abdominal CT. , 2000, Radiology.

[9]  H. Kauczor,et al.  Dual-Energy Computed-Tomography Cholangiography in Potential Donors for Living-Related Liver Transplantation: Initial Experience , 2010, Investigative radiology.

[10]  Bernhard Schmidt,et al.  Dual Source CT Technology , 2008 .

[11]  D. Hough,et al.  Dual-energy and dual-source CT: is there a role in the abdomen and pelvis? , 2009, Radiologic clinics of North America.

[12]  M. Rengo,et al.  Dual energy CT (DECT) of the liver: conventional versus virtual unenhanced images , 2010, European Radiology.

[13]  J. Paul,et al.  Low-kilovoltage multi-detector row chest CT in adults: feasibility and effect on image quality and iodine dose. , 2004, Radiology.

[14]  P. Seidensticker,et al.  Dual source CT imaging , 2008 .

[15]  Hersh Chandarana,et al.  Dual-source dual-energy MDCT of pancreatic adenocarcinoma: initial observations with data generated at 80 kVp and at simulated weighted-average 120 kVp. , 2010, AJR. American journal of roentgenology.

[16]  Bernhard Schmidt,et al.  Radiation dose and image quality in pediatric CT: effect of technical factors and phantom size and shape. , 2004, Radiology.

[17]  Simon Wildermuth,et al.  Low kilovoltage cardiac dual-source CT: attenuation, noise, and radiation dose , 2008, European Radiology.

[18]  Jin Peng,et al.  Liver virtual non-enhanced CT with dual-source, dual-energy CT: a preliminary study , 2010, European Radiology.

[19]  M. Macari,et al.  Dual energy CT: preliminary observations and potential clinical applications in the abdomen , 2008, European Radiology.

[20]  M. Reiser,et al.  Material differentiation by dual energy CT: initial experience , 2007, European Radiology.

[21]  Multidetector Computed Tomography Chest Examinations With Low-Kilovoltage Protocols in Adults: Effect on Image Quality and Radiation Dose , 2009, Journal of computer assisted tomography.

[22]  Ehsan Samei,et al.  Hypervascular liver tumors: low tube voltage, high tube current multidetector CT during late hepatic arterial phase for detection--initial clinical experience. , 2009, Radiology.

[23]  Hersh Chandarana,et al.  Abdominal aorta: evaluation with dual-source dual-energy multidetector CT after endovascular repair of aneurysms--initial observations. , 2008, Radiology.

[24]  Jae Hyung Park,et al.  The Image Quality and Radiation Dose of 100-kVp versus 120-kVp ECG-Gated 16-Slice CT Coronary Angiography , 2009, Korean journal of radiology.

[25]  S. Schindera,et al.  Thoracoabdominal-Aortoiliac Multidetector-Row CT Angiography at 80 and 100 kVp: Assessment of Image Quality and Radiation Dose , 2009, Investigative radiology.