Round robin test on a cavity resonance method to measure complex permittivity of dielectric plates at microwave frequency [Errata]

A round robin test on a cavity resonance method was performed by seven groups in Japan, where a cylindrical cavity is cut into two halves at the middle of its length and a dielectric plate sample is placed between these two halves. On the basis of the rigorous analysis by the Ritz-Galerkin method to take account of fringing field in the plate region outside diameter of the cavity, the relative permittivity epsiv' and loss tangent tandelta are determined accurately from measured resonant frequency and unloaded Q for the TE011 mode. The measured results are presented for epsiv', tandelta, temperature coefficient of relative permittivity TCepsiv, and that of loss tangent TCtandelta. Measurement accuracies estimated by considering the measurement error in each group and the amount of scatter in measured values among these groups are 0.3% for epsiv' and 4% for tandelta of 10-4 and 20% for 10-6. The measurement resolution of TCepsiv and TCtandelta are estimated to be 1 ppm/K and 0.04 %/K for (Zr,Sn)TiO4 ceramics with TCepsiv=-10.0 ppm/K and TCtandelta=0.24%/K. The measurement error Deltaepsiv'/epsiv' by the cavity resonance method is about half of ones by the dielectric rod resonance method. This high measurement accuracy and resolution are acceptable for most practical applications to microwave planar circuits

[1]  Y. Kobayashi,et al.  Microwave Measurement of Dielectric Properties of Low-Loss Materials by the Dielectric Rod Resonator Method , 1985 .

[2]  G. Kent An evanescent-mode tester for ceramic dielectric substrates , 1988 .

[3]  Yoshio Kobayashi,et al.  Cut-Off Circular Waveguide Method for Dielectric Substrate Measurements in Millimeter Wave Range(General Methods, Materials, and Passive Circuits)( Advances in Characterization and Measurement Technologies for Microwave and Millimeter-Wave Materials, Devices and Circuits) , 2004 .

[4]  Y. Kobayashi,et al.  Improved Cavity Resonance Method For Nondestructive Measurement Of Complex Permittivity Of Dielectric Plate , 1988, 1988 Conference on Precision Electromagnetic Measurements.

[5]  B. W. Hakki,et al.  A Dielectric Resonator Method of Measuring Inductive Capacities in the Millimeter Range , 1960 .

[6]  Yoshio Kobayashi,et al.  Round Robin Test on a Dielectric Resonator Method for Measuring Complex Permittivity at Microwave Frequency (Special Issue on Measurement Techniques for Microwave/Millimeter Wave) , 1994 .

[7]  Graham Kent Non-destructive permittivity measurement of substrates , 1994 .

[8]  G. Kent,et al.  The gap correction for the resonant-mode dielectrometer , 1996 .

[9]  Y. Kobayashi,et al.  Round robin test on a cavity resonance method to measure complex permittivity of dielectric plates at microwave frequency , 2006, IEEE Transactions on Dielectrics and Electrical Insulation.

[10]  Yoshio Kobayashi,et al.  Complex Permittivity Measurement of Dielectric Plates by a Cavity Resonance Method , 1988 .