Monte-Carlo simulation of the Siemens Artiste linear accelerator flat 6 MV and flattening-filter-free 7 MV beam line

The aim of our work is to provide the up-to-now missing information on the Siemens Artiste FFF 7 MV beam line using a Monte-Carlo model fit to the realistic dosimetric measurements at the linear accelerator in clinical use at our department. The main Siemens Artiste 6MV and FFF 7MV beams were simulated using the Geant4 toolkit. The simulations were compared with the measurements with an ionization chamber in a water phantom to verify the validation of simulation and tuning the primary electron parameters. Hereafter, other parameters such as surface dose, spectrum, electron contamination, symmetry, flatness/unflatness, slope, and characteristic off-axis changes were discussed for both Flat and FFF mode. The mean electron energy for the FFF beam was 8.8 MeV and 7.5 MeV for Flat 6 MV, the spread energy and spot size of the selected Gaussian distribution source were 0.4 MeV and 1mm, respectively. The dose rate of the FFF beam was 2.8 (2.96) times higher than for the flattened beam for a field size of 10×10 (20×20) cm2. The electron contamination has significant contribution to the surface dose especially for the flattened beam. The penumbra, surface dose and the mean energy of photons decrease by removing the flattening filter. Finally, the results show that off-axis changes have no strong effect on the mean energy of FFF beams, while this effect was more considerable for the flattened beam.

[1]  H. Dowlatabadi,et al.  Benchmarking of Siemens Linac in Electron Modes: 8-14 MeV Electron Beams , 2018, Journal of biomedical physics & engineering.

[2]  E. Chakir,et al.  Evaluation of the dosimetric characteristics of 6 MV flattened and unflattened photon beam , 2017 .

[3]  C. Sureka,et al.  Comparison of Flattening Filter (FF) and Flattening-Filter-Free (FFF) 6 MV photon beam characteristics for small field dosimetry using EGSnrc Monte Carlo code , 2017 .

[4]  Klemens Zink,et al.  Monte Carlo‐based investigations on the impact of removing the flattening filter on beam quality specifiers for photon beam dosimetry , 2017, Medical physics.

[5]  Arunai Nambiraj,et al.  Comparison of surface dose delivered by 7 MV-unflattened and 6 MV-flattened photon beams. , 2017, Reports of practical oncology and radiotherapy : journal of Greatpoland Cancer Center in Poznan and Polish Society of Radiation Oncology.

[6]  Neeraj Sharma,et al.  Monte Carlo study of a flattening filter-free 6 MV photon beam using the BEAMnrc code , 2017 .

[7]  C. Rübe,et al.  mARC Treatment of Hypopharynx Carcinoma with Flat and Flattening-Filter-Free Beam Energies – A Planning Study , 2016, PloS one.

[8]  C. Rübe,et al.  mARC prostate treatment planning with Varian Eclipse for flat vs. FFF beams. , 2016, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[9]  B. Paliwal,et al.  Static IMRT and VMAT planning on 6 MV Flattened and Flattening-Filter-Free Beams of a TrueBeam VirtuaLinac System , 2015, 1509.06845.

[10]  Niko Papanikolaou,et al.  Flattening filter‐free accelerators: a report from the AAPM Therapy Emerging Technology Assessment Work Group , 2015, Journal of applied clinical medical physics.

[11]  Long Huang,et al.  Planning for mARC treatments with the Eclipse treatment planning system , 2015, Journal of applied clinical medical physics.

[12]  F. Seif,et al.  Evaluation of Electron Contamination in Cancer Treatment with Megavoltage Photon Beams: Monte Carlo Study , 2015, Journal of Biomedical Physics and Engineering.

[13]  A. Ghasemi,et al.  Determination of primary electron beam parameters in a Siemens Primus Linac using Monte Carlo simulation , 2015 .

[14]  C. Rübe,et al.  mARC vs. IMRT radiotherapy of the prostate with flat and flattening-filter-free beam energies , 2014, Radiation oncology.

[15]  N. Khelassi-Toutaoui,et al.  Monte Carlo photon beam modeling and commissioning for radiotherapy dose calculation algorithm. , 2014, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[16]  K. Balaji,et al.  Analysis of physical parameters and determination of inflection point for Flattening Filter Free beams in medical linear accelerator. , 2014, Reports of practical oncology and radiotherapy : journal of Greatpoland Cancer Center in Poznan and Polish Society of Radiation Oncology.

[17]  C. Rübe,et al.  Comparative Planning of Flattening-Filter-Free and Flat Beam IMRT for Hypopharynx Cancer as a Function of Beam and Segment Number , 2014, PloS one.

[18]  Jan Palm,et al.  Planning study and dose measurements of intracranial stereotactic radiation surgery with a flattening filter-free linac. , 2014, Practical radiation oncology.

[19]  Mehran Yarahmadi,et al.  Improvement of the penumbra for small radiosurgical fields using flattening filter free low megavoltage beams. , 2013, Zeitschrift fur medizinische Physik.

[20]  Y. Yoshioka,et al.  Characteristics of flattening filter free beams at low monitor unit settings. , 2013, Medical physics.

[21]  Keyvan Jabbari,et al.  Monte Carlo Simulation of Siemens ONCOR Linear Accelerator with BEAMnrc and DOSXYZnrc Code , 2013, Journal of medical signals and sensors.

[22]  Arne Blechschmidt,et al.  The impact of flattening-filter-free beam technology on 3D conformal RT , 2013, Radiation Oncology.

[23]  Yvonne Dzierma,et al.  Visualisation of Respiratory Tumour Motion and Co-Moving Isodose Lines in the Context of Respiratory Gating, IMRT and Flattening-Filter-Free Beams , 2013, PloS one.

[24]  M. Özdal,et al.  The investigation of cardiac functions and blood lipid levels in obese men and elite athletes. , 2013 .

[25]  A. Zakaria,et al.  Monte Carlo Modeling of a Siemens Primus 6 MV Photon Beam Linear Accelerator , 2013 .

[26]  Joseph Perl,et al.  The Geant4 Visualisation System - a multi-driver graphics system , 2012, ArXiv.

[27]  A Fogliata,et al.  Definition of parameters for quality assurance of flattening filter free (FFF) photon beams in radiation therapy. , 2012, Medical physics.

[28]  R. Siochi,et al.  Dosimetric properties of a beam quality‐matched 6 MV unflattened photon beam , 2012, Journal of applied clinical medical physics.

[29]  Yvonne Dzierma,et al.  Beam properties and stability of a flattening-filter free 7 MV beam-an overview. , 2012, Medical physics.

[30]  H. Shukla,et al.  A planning and delivery study of a rotational IMRT technique with burst delivery. , 2011, Medical physics.

[31]  B. Poppe,et al.  A direction-selective flattening filter for clinical photon beams. Monte Carlo evaluation of a new concept , 2011, Physics in medicine and biology.

[32]  D. Visvikis,et al.  SU‐E‐T‐715: GATE Monte Carlo Simulation of a 6 MV SIEMENS OncorTM Linear Accelerator Photon Beam : Feasibility and Preliminary Dosimetric Study , 2011 .

[33]  Elinore Wieslander,et al.  A Monte Carlo study of a flattening filter-free linear accelerator verified with measurements , 2010, Physics in medicine and biology.

[34]  Asghar Mesbahi,et al.  Dosimetric characteristics of unflattened 6 MV photon beams of a clinical linear accelerator: a Monte Carlo study. , 2007, Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine.

[35]  Ahmad Keshtkar,et al.  Dosimetric properties of a flattening filter-free 6-MV photon beam: a Monte Carlo study , 2007, Radiation Medicine.

[36]  M. Alber,et al.  A virtual photon source model of an Elekta linear accelerator with integrated mini MLC for Monte Carlo based IMRT dose calculation , 2007, Physics in medicine and biology.

[37]  B Serrano,et al.  Monte Carlo simulation of a medical linear accelerator for radiotherapy use. , 2006, Radiation protection dosimetry.

[38]  Radhe Mohan,et al.  Properties of unflattened photon beams shaped by a multileaf collimator. , 2006, Medical physics.

[39]  Radhe Mohan,et al.  Monte Carlo study of photon fields from a flattening filter-free clinical accelerator. , 2006, Medical physics.

[40]  J Yang,et al.  Modelling of electron contamination in clinical photon beams for Monte Carlo dose calculation. , 2004, Physics in medicine and biology.

[41]  Frank Verhaegen,et al.  Monte Carlo modelling of external radiotherapy photon beams. , 2003, Physics in medicine and biology.

[42]  A. Dell'Acqua,et al.  Geant4 - A simulation toolkit , 2003 .

[43]  I. Kawrakow,et al.  History by history statistical estimators in the BEAM code system. , 2002, Medical physics.

[44]  I. Kawrakow Accurate condensed history Monte Carlo simulation of electron transport. II. Application to ion chamber response simulations. , 2000, Medical physics.

[45]  A Kapur,et al.  Electron beam modeling and commissioning for Monte Carlo treatment planning. , 2000, Medical physics.

[46]  T. Mackie,et al.  A dual source photon beam model used in convolution/superposition dose calculations for clinical megavoltage x-ray beams. , 1997, Medical physics.