Wear analysis of chamfered elongated acetabular cup liners

This study investigated the wear phenomena of chamfered acetabular cup liners. The liners have three parameters at the lengthened rim: the length of the elongation, the depth of the cup and the chamfer angle. Using published wear volume equations for cylindrically elongated liners, this study analysed the volume of the chamfer to obtain the exact theoretical wear volume of the chamfered liner. The criteria described in our previous paper were used to verify the accuracy of the proposed methodology. The results showed that a shallow cup depth (<2 mm) and a chamfer angle of about 30° could significantly reduce the wear volume arising from the elongation. When the chamfer angle was increased further, this effect became progressively less significant. The results suggest that a chamfer angle up to about 30° is appropriate, but the angle should not be larger than required to obtain the minimum femoral range of motion (ROM). The results of this study can also be integrated with the analysis of the kinetics and kinematics of liners. A compromise design incorporating a lower wear volume, sufficient ROM and a lower contact stress should be possible.

[1]  The effect of component design on range of motion to impingement in total hip arthroplasty. , 2001, Instructional course lectures.

[2]  M H Aliabadi,et al.  Total hip arthroplasty wear simulation using the boundary element method. , 2007, Journal of biomechanics.

[3]  M. Conditt,et al.  Does neck/liner impingement increase wear of ultrahigh-molecular-weight polyethylene liners? , 2006, The Journal of arthroplasty.

[4]  B. Masri,et al.  Treatment of hip instability. , 2001, The Orthopedic clinics of North America.

[5]  R. Brumback,et al.  Functional evaluation of the shoulder after transfer of the vascularized latissimus dorsi muscle. , 1992, The Journal of bone and joint surgery. American volume.

[6]  S H Teoh,et al.  An elasto-plastic finite element model for polyethylene wear in total hip arthroplasty. , 2002, Journal of biomechanics.

[7]  I. Clarke,et al.  How do alternative bearing surfaces influence wear behavior? , 2008, The Journal of the American Academy of Orthopaedic Surgeons.

[8]  James Shih-Shyn Wu,et al.  Evaluating the accuracy of wear formulae for acetabular cup liners , 2010, Medical & Biological Engineering & Computing.

[9]  A. Edidin,et al.  A literature review of the association between wear rate and osteolysis in total hip arthroplasty. , 2002, The Journal of arthroplasty.

[10]  C. Engh,et al.  Comparison of in vivo wear between polyethylene liners articulating with ceramic and cobalt-chrome femoral heads , 2000 .

[11]  R. Barrack,et al.  Dislocation After Total Hip Arthroplasty: Implant Design and Orientation , 2003, The Journal of the American Academy of Orthopaedic Surgeons.

[12]  James Shih-Shyn Wu,et al.  Wear patterns of, and wear volume formulae for, cylindrically elongated acetabular cup liners , 2010, Medical & Biological Engineering & Computing.

[13]  E. Garcia-Cimbrelo,et al.  Polyethylene in total hip arthroplasty: half a century in the limelight , 2010, Journal of Orthopaedics and Traumatology.

[14]  D. Dowson,et al.  An Investigation into the Origins of Time-Dependent Variation in Penetration Rates with Charnley Acetabular Cups—Wear, Creep or Degradation? , 1996, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[15]  J F Orr,et al.  Wear paths produced by individual hip-replacement patients--a large-scale, long-term follow-up study. , 2008, Journal of biomechanics.

[16]  J. Charnley,et al.  The optimum size of prosthetic heads in relation to the wear of plastic sockets in total replacement of the hip , 2006, Medical and biological engineering.

[17]  A. Iglič,et al.  Polyethylene wear in total hip prostheses: the influence of direction of linear wear on volumetric wear determined from radiographic data , 2003, Skeletal Radiology.

[18]  Kengo Yamamoto,et al.  Validation of acetabular cup wear volume based on direct and two-dimensional measurements: hip simulator analysis , 2003, Journal of orthopaedic science : official journal of the Japanese Orthopaedic Association.

[19]  B. Derbyshire The estimation of acetabular cup wear volume from two-dimensional measurements: A comprehensive analysis , 1998, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[20]  B. Morrey,et al.  Effect of femoral head size on wear of the polyethylene acetabular component. , 1990, The Journal of bone and joint surgery. American volume.

[21]  T. Wright,et al.  Dislocation following THA: comparison of two acetabular component designs. , 1993, Orthopedics.

[22]  M. Ritter,et al.  Total hip arthroplasty in patients with osteonecrosis. The effect of cement techniques. , 1997, Clinical orthopaedics and related research.

[23]  C. Shih,et al.  MEASUREMENT OF POLYETHYLENE WEAR IN CEMENTLESS TOTAL HIP ARTHROPLASTY , 1997 .

[24]  S. Banks,et al.  Association between dislocation, impingement, and articular geometry in retrieved acetabular polyethylene cups , 2007, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[25]  P. Devane,et al.  Three-dimensional polyethylene wear of a press-fit titanium prosthesis. Factors influencing generation of polyethylene debris. , 1997, The Journal of arthroplasty.

[26]  James Shih-Shyn Wu,et al.  Wear patterns of, and wear volume formulae for, hemispherical acetabular cup liners , 2010 .

[27]  B. Wroblewski,et al.  Cemented revision for aseptic acetabular loosening. A review of 387 hips. , 1995, The Journal of bone and joint surgery. British volume.

[28]  M. Reimold,et al.  Estimation of the wear volume after total hip replacement. A simple access to geometrical concepts. , 2008, Medical engineering & physics.

[29]  James Shih-Shyn Wu,et al.  The computer simulation of wear behavior appearing in total hip prosthesis , 2003, Comput. Methods Programs Biomed..

[30]  R B Bourne,et al.  Measurement of polyethylene wear in metal-backed acetabular cups. I. Three-dimensional technique. , 1995, Clinical orthopaedics and related research.

[31]  John Fisher,et al.  The role of macrophages in osteolysis of total joint replacement. , 2005, Biomaterials.

[32]  James Shih-Shyn Wu,et al.  Measurement of polyethylene wear - a new three-dimensional methodology , 2002, Comput. Methods Programs Biomed..

[33]  W. Harris,et al.  The problem is osteolysis. , 1995, Clinical orthopaedics and related research.

[34]  T. Bauer,et al.  Three-Dimensional Analysis of Multiple Wear Vectors in Retrieved Acetabular Cups* , 1997, The Journal of bone and joint surgery. American volume.

[35]  Daniel Kluess,et al.  Influence of femoral head size on impingement, dislocation and stress distribution in total hip replacement. , 2007, Medical engineering & physics.

[36]  B. Morrey,et al.  Effect of the Elevated-Rim Acetabular Liner on Loosening after Total Hip Arthroplasty* , 1997, The Journal of bone and joint surgery. American volume.

[37]  P. Rullkoetter,et al.  Finite element simulation of early creep and wear in total hip arthroplasty. , 2005, Journal of biomechanics.

[38]  J. Hung,et al.  A COMPARATIVE STUDY ON WEAR BEHAVIOR OF HIP PROSTHESIS BY FINITE ELEMENT SIMULATION , 2002 .

[39]  D. Sochart,et al.  Relationship of acetabular wear to osteolysis and loosening in total hip arthroplasty. , 1999, Clinical orthopaedics and related research.

[40]  D. Murray Impingement and loosening of the long posterior wall acetabular implant. , 1992, The Journal of bone and joint surgery. British volume.

[41]  C. Shih,et al.  Measurement of polyethylene wear in cementless total hip arthroplasty. , 1997, The Journal of bone and joint surgery. British volume.

[42]  D J Berry,et al.  Epidemiology of instability after total hip replacement. , 2001, The Orthopedic clinics of North America.

[43]  Aditya Maheshwari,et al.  Impingement with total hip replacement. , 2007, The Journal of bone and joint surgery. American volume.

[44]  S. Leopold,et al.  The effect of joint loading on acetabular wear measurement in total hip arthroplasty. , 2000, The Journal of arthroplasty.