Theoretical and experimental analysis of air cooling for intracavitary microwave hyperthermia applicators

An intracavitary microwave antenna array system has been developed and tested for the hyperthermia treatment of prostate cancer at Thayer School of Engineering and Dartmouth-Witchcock Medical Center. The antenna array consists of a choked dipole antenna inserted into the urethra and a choked dipole antenna eccentrically embedded in a Teflon obturator inserted into the rectum. To prevent unnecessary heating of the healthy tissue that surrounds each applicator, an air cooling system has been incorporated into the rectal applicator. The air cooling system was designed and modeled theoretically using a numerical solution of heat and momentum equations within the applicator, and an analytical solution of the Pennes bioheat equation in tissue surrounding the applicator. The 3D temperature distribution produced by the air-cooled rectal applicator was measured in a perfused canine prostate.<<ETX>>

[1]  Electromagnetic and thermal models of a water-cooled dipole radiating in a biological tissue , 1991, IEEE Transactions on Biomedical Engineering.

[2]  The effect of air cooling on the radial temperature distribution of a single microwave hyperthermia antenna in vivo. , 1991, International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group.

[3]  E. Jonsson,et al.  A coaxial microwave applicator for transurethral hyperthermia of the prostate , 1993, The Prostate.

[4]  V. Eppert,et al.  Air cooling for an interstitial microwave hyperthermia antenna: theory and experiment , 1991, IEEE Transactions on Biomedical Engineering.

[5]  J W Strohbehn,et al.  Some aspects of optimization of an invasive microwave antenna for local hyperthermia treatment of cancer. , 1981, Medical physics.