Dynamic coherence factor based on the standard deviation for coherent plane-wave compounding

The ultrafast imaging technique based on plane wave transmission has been a commonly investigated imaging mode in medical ultrasound imaging. Coherent plane-wave compounding (CPWC) was proposed to improve the quality of plane-wave imaging (PWI), which obtained a high-quality image by summing the low-quality images formed by transmitting plane waves at different steering angles. Coherence factor (CF) weighting algorithms can effectively improve image contrast with low computational complexity. However, this usually introduces black artifacts and degraded speckle quality. In this paper, we propose a dynamic coherence factor (DCF) that is based on the angular difference of CPWC to adaptively determine the number of plane waves and utilizes several plane waves with small angular difference to evaluate coherence. To improve resolution and contrast-to-noise ratio (CNR), adjusted DCF (ADCF) introduces a parameter related to the standard deviation to adjust DCF. Furthermore, a square neighborhood ADCF (SN-ADCF) using a 2-dimensional average filter is designed to obtain higher contrast and speckle quality. The simulated, experimental and in-vivo datasets are used to evaluate the proposed methods. Results show that, in comparison with CF, DCF can achieve improved contrast ratio (CR), CNR and speckle signal-to-noise ratio (sSNR). ADCF achieves a maximal lateral full width at half maximum (FWHM) improvement by 40% and better speckle quality than CF. SN-ADCF causes about 20% improvements upon CF in CNR and sSNR, while maintaining a similar resolution performance. SN-ADCF also provides higher lateral resolution than the generalized coherence factor (GCF) and scaled coherence factor (scCF), meanwhile obtaining a comparable contrast and speckle quality. Therefore, SN-ADCF has a satisfying comprehensive performance, which can achieve a reasonable balance among resolution, contrast and speckle quality.

[1]  Sheng-Min Huang,et al.  Beamforming effects on generalized Nakagami imaging , 2015, Physics in medicine and biology.

[2]  Hu Peng,et al.  Coherent Plane-Wave Compounding Based on Normalized Autocorrelation Factor , 2018, IEEE Access.

[3]  M. Fink,et al.  Ultrasound contrast plane wave imaging , 2012, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control.

[4]  Yong-Ping Zheng,et al.  Subarray coherence based postfilter for eigenspace based minimum variance beamformer in ultrasound plane-wave imaging. , 2016, Ultrasonics.

[5]  Bin Du,et al.  Coherence Plane-Wave Compounding with Angle Coherence Factor for Ultrafast Ultrasound Imaging , 2018, 2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC).

[6]  Pai-Chi Li,et al.  SNR-dependent coherence-based adaptive imaging for high-frame-rate ultrasonic and photoacoustic imaging , 2014, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control.

[7]  Mathieu Pernot,et al.  Ultrafast Doppler Imaging of Blood Flow Dynamics in the Myocardium , 2012, IEEE Transactions on Medical Imaging.

[8]  Shun Zhang,et al.  Short-lag Spatial Coherence Ultrasound Imaging with Adaptive Synthetic Transmit Aperture Focusing , 2017, Ultrasonic imaging.

[9]  Sverre Holm,et al.  Wiener beamforming and the coherence factor in ultrasound imaging , 2010, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[10]  Xin Yang,et al.  Spatio-temporally smoothed coherence factor for ultrasound imaging [Correspondence] , 2014, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control.

[12]  Jian-yu Lu,et al.  Extended high-frame rate imaging method with limited-diffraction beams , 2006, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[13]  J. Lu,et al.  Experimental study of high frame rate imaging with limited diffraction beams. , 1998, IEEE transactions on ultrasonics, ferroelectrics, and frequency control.

[14]  Armando Manduca,et al.  Comb-Push Ultrasound Shear Elastography (CUSE) With Various Ultrasound Push Beams , 2013, IEEE Transactions on Medical Imaging.

[15]  Moein Mozaffarzadeh,et al.  Enhanced linear-array photoacoustic beamforming using modified coherence factor , 2017, Journal of biomedical optics.

[16]  Mickael Tanter,et al.  Ultrafast imaging in biomedical ultrasound , 2014 .

[17]  Jian-yu Lu 2D and 3D high frame rate imaging with limited diffraction beams , 1997 .

[18]  K. Boone,et al.  Effect of skin impedance on image quality and variability in electrical impedance tomography: a model study , 1996, Medical and Biological Engineering and Computing.

[19]  J. Camacho,et al.  Phase Coherence Imaging , 2009, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[20]  Yuanyuan Wang,et al.  Ultrasound harmonic enhanced imaging using eigenspace-based coherence factor. , 2016, Ultrasonics.

[21]  Wei Guo,et al.  Joint Subarray Coherence and Minimum Variance Beamformer for Multitransmission Ultrasound Imaging Modalities , 2018, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control.

[22]  Xun Chen,et al.  Short-lag spatial coherence combined with eigenspace-based minimum variance beamformer for synthetic aperture ultrasound imaging , 2017, Comput. Biol. Medicine.

[23]  M. Fink,et al.  Coherent plane-wave compounding for very high frame rate ultrasonography and transient elastography , 2009, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[24]  J. Jensen,et al.  Calculation of pressure fields from arbitrarily shaped, apodized, and excited ultrasound transducers , 1992, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[25]  Yuanyuan Wang,et al.  Eigenspace-based minimum variance beamformer combined with Wiener postfilter for medical ultrasound imaging. , 2012, Ultrasonics.

[26]  Ole Marius Hoel Rindal,et al.  The influence of speckle statistics on contrast metrics in ultrasound imaging , 2017, 2017 IEEE International Ultrasonics Symposium (IUS).

[27]  Jeremy J Dahl,et al.  Angular coherence in ultrasound imaging: Theory and applications. , 2017, The Journal of the Acoustical Society of America.

[28]  Pai-Chi Li,et al.  Adaptive imaging using the generalized coherence factor. , 2003, IEEE transactions on ultrasonics, ferroelectrics, and frequency control.

[29]  F. Gran,et al.  Broadband minimum variance beamforming for ultrasound imaging , 2009, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[30]  Hu Peng,et al.  An adaptive beamforming method for ultrasound imaging based on the mean‐to‐standard‐deviation factor , 2018, Ultrasonics.

[31]  Yang Zhang,et al.  Ultrafast Ultrasound Imaging Using Combined Transmissions With Cross-Coherence-Based Reconstruction , 2018, IEEE Transactions on Medical Imaging.

[32]  Armando Manduca,et al.  Comb-Push Ultrasound Shear Elastography (CUSE): A Novel Method for Two-Dimensional Shear Elasticity Imaging of Soft Tissues , 2012, IEEE Transactions on Medical Imaging.

[33]  G. E. Trahey,et al.  Short-lag spatial coherence of backscattered echoes: imaging characteristics , 2011, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[34]  Wei Guo,et al.  Sidelobe reduction for plane wave compounding with a limited frame number , 2018, Biomedical engineering online.

[35]  He Li,et al.  Systematic Performance Evaluation of a Cross-Correlation-Based Ultrasound Strain Imaging Method. , 2016, Ultrasound in medicine & biology.

[36]  Tai-Kyong Song,et al.  Methods for Grating Lobe Suppression in Ultrasound Plane Wave Imaging , 2018, Applied Sciences.