PURPOSE Accurate tumor tracking remains a challenge in current radiation therapy. Many strategies including image guided radiation therapy alleviate the problem to certain extents. The authors propose a new modality called emission guided radiation therapy (EGRT) to accurately and directly track the tumor based on its biological signature. This work is to demonstrate the feasibility of EGRT under two clinical scenarios using a 4D digital patient model. METHODS EGRT uses lines of response (LOR's) from positron emission events to direct beamlets of therapeutic radiation through the emission sites inside a tumor. This is accomplished by a radiation delivery system consisting of a Linac and positron emission tomography (PET) detectors on a fast rotating closed-ring gantry. During the treatment of radiotracer-administrated cancer patients, PET detectors collect LOR's from tumor uptake sites and the Linac responds in nearly real-time with beamlets of radiation along the same LOR paths. Moving tumors are therefore treated with a high targeting accuracy. Based on the EGRT concept, the authors design a treatment method with additional modulation algorithms including attenuation correction and an integrated boost scheme. Performance is evaluated using simulations of a lung tumor case with 3D motion and a prostate tumor case with setup errors. The emission process is simulated by Geant4 Application for Tomographic Emission package (GATE) and Linac dose delivery is simulated using a voxel-based Monte Carlo algorithm (VMC++). RESULTS In the lung case with attenuation correction, compared to a conventional helical treatment, EGRT achieves a 41% relative increase in dose to 95% of the gross tumor volume (GTV) and a 55% increase to 50% of the GTV. All dose distributions are normalized for the same dose to the lung. In the prostate case with the integrated boost and no setup error, EGRT yields a 19% and 55% relative dose increase to 95% and 50% of the GTV, respectively, when all methods are normalized for the same dose to the rectum. In the prostate case with integrated boost where setup error is present, EGRT contributes a 21% and 52% relative dose increase to 95% and 50% of the GTV, respectively. CONCLUSIONS As a new radiation therapy modality with inherent tumor tracking, EGRT has the potential to substantially improve targeting in radiation therapy in the presence of intrafractional and interfractional motion.

[1]  K. Lam,et al.  Retrospective analysis of prostate cancer patients with implanted gold markers using off-line and adaptive therapy protocols. , 2005, International journal of radiation oncology, biology, physics.

[2]  W P Segars,et al.  Realistic CT simulation using the 4D XCAT phantom. , 2008, Medical physics.

[3]  I. Rosen,et al.  Implementing IMRT in clinical practice: a joint document of the American Society for Therapeutic Radiology and Oncology and the American Association of Physicists in Medicine. , 2004, International journal of radiation oncology, biology, physics.

[4]  Thomas Beyer,et al.  Radiation exposure of patients undergoing whole-body dual-modality 18F-FDG PET/CT examinations. , 2005, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[5]  S Webb,et al.  IMRT delivery to a moving target by dynamic MLC tracking: delivery for targets moving in two dimensions in the beam's eye view , 2006, Physics in medicine and biology.

[6]  Steve B. Jiang,et al.  Fluoroscopic tumor tracking for image-guided lung cancer radiotherapy , 2009, Physics in medicine and biology.

[7]  M. Petasecca,et al.  Independent quality assurance of a helical tomotherapy machine using the dose magnifying glass. , 2011, Medical Physics (Lancaster).

[8]  Lei Xing,et al.  Total-Variation Regularization Based Inverse Planning for Intensity Modulated Arc Therapy , 2012, Technology in cancer research & treatment.

[9]  Herbert Cattell,et al.  Geometric accuracy of a real-time target tracking system with dynamic multileaf collimator tracking system. , 2006, International journal of radiation oncology, biology, physics.

[10]  Jean-Philippe Pignol,et al.  Correlation of lung tumor motion with external surrogate indicators of respiration. , 2004, International journal of radiation oncology, biology, physics.

[11]  R. Siddon Fast calculation of the exact radiological path for a three-dimensional CT array. , 1985, Medical physics.

[12]  J A Purdy,et al.  Clinical dose-volume histogram analysis for pneumonitis after 3D treatment for non-small cell lung cancer (NSCLC) , 1999, International journal of radiation oncology, biology, physics.

[13]  Daniela Thorwarth,et al.  Physical radiotherapy treatment planning based on functional PET/CT data. , 2010, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[14]  Steve B. Jiang,et al.  Towards fluoroscopic respiratory gating for lung tumours without radiopaque markers , 2005, Physics in medicine and biology.

[15]  I. Kawrakow,et al.  3D electron dose calculation using a Voxel based Monte Carlo algorithm (VMC). , 1996, Medical physics.

[16]  T. Landberg,et al.  What margins should be added to the clinical target volume in radiotherapy treatment planning for lung cancer? , 1998, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[17]  F DuBois Bowman,et al.  Initial experience with the radiotracer anti-1-amino-3-18F-fluorocyclobutane-1-carboxylic acid with PET/CT in prostate carcinoma. , 2007, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[18]  S. Nehmeh,et al.  Validation of GATE Monte Carlo simulations of the GE Advance/Discovery LS PET scanners. , 2005, Medical physics.

[19]  Yoshihiro Takai,et al.  Evaluation of inter- and intrafraction organ motion during intensity modulated radiation therapy (IMRT) for localized prostate cancer measured by a newly developed on-board image-guided system. , 2005, Radiation medicine.

[20]  Paul Kinahan,et al.  A combined PET/CT scanner for clinical oncology. , 2000, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[21]  Lei Xing,et al.  Using total-variation regularization for intensity modulated radiation therapy inverse planning with field-specific numbers of segments , 2008, Physics in medicine and biology.

[22]  T. Turkington,et al.  FDG imaging of lung nodules: a phantom study comparing SPECT, camera-based PET, and dedicated PET. , 1999, Radiology.

[23]  Daniel Y Sze,et al.  CT-guided transthoracic needle aspiration biopsy of pulmonary nodules: needle size and pneumothorax rate. , 2003, Radiology.

[24]  Paul J. Reckwerdt,et al.  Tomotherapy: Optimized planning and delivery of radiation therapy , 1995, Int. J. Imaging Syst. Technol..

[25]  Hazim Jaradat,et al.  The utility of megavoltage computed tomography images from a helical tomotherapy system for setup verification purposes. , 2004, International journal of radiation oncology, biology, physics.

[26]  Y. Tsunashima,et al.  Correlation between the respiratory waveform measured using a respiratory sensor and 3D tumor motion in gated radiotherapy. , 2004, International journal of radiation oncology, biology, physics.

[27]  G. V. von Schulthess,et al.  Fluorocholine PET/CT in patients with prostate cancer: initial experience. , 2005, Radiology.

[28]  Niko Papanikolaou,et al.  QA for helical tomotherapy: Report of the AAPM Task Group 148a). , 2010, Medical physics.

[29]  R. Mohan,et al.  Simultaneous integrated boost intensity-modulated radiotherapy for locally advanced head-and-neck squamous cell carcinomas. I: dosimetric results. , 2003, International journal of radiation oncology, biology, physics.

[30]  Steve B. Jiang,et al.  Multiple template-based fluoroscopic tracking of lung tumor mass without implanted fiducial markers , 2007, Physics in medicine and biology.

[31]  I. Kawrakow VMC++, Electron and Photon Monte Carlo Calculations Optimized for Radiation Treatment Planning , 2001 .

[32]  Martin J Murphy,et al.  Tracking moving organs in real time. , 2004, Seminars in radiation oncology.

[33]  Sergey Povzner,et al.  Management of three-dimensional intrafraction motion through real-time DMLC tracking. , 2008, Medical physics.

[34]  C Lartizien,et al.  GATE: a simulation toolkit for PET and SPECT. , 2004, Physics in medicine and biology.

[35]  Vincent Breton,et al.  GATE (geant4 application for tomographic emission): a PET/SPECT general-purpose simulation platform , 2003 .

[36]  H. Kubo,et al.  Respiration gated radiotherapy treatment: a technical study. , 1996, Physics in medicine and biology.

[37]  Steven D Chang,et al.  An Analysis of the Accuracy of the CyberKnife: A Robotic Frameless Stereotactic Radiosurgical System , 2003, Neurosurgery.

[38]  C. Yiannoutsos,et al.  A pilot trial of serial 18F-fluorodeoxyglucose positron emission tomography in patients with medically inoperable stage I non-small-cell lung cancer treated with hypofractionated stereotactic body radiotherapy. , 2010, International journal of radiation oncology, biology, physics.

[39]  D Verellen,et al.  A computerized remote table control for fast on-line patient repositioning: implementation and clinical feasibility. , 2000, Medical physics.

[40]  Matthias Guckenberger,et al.  Intra-fractional uncertainties in cone-beam CT based image-guided radiotherapy (IGRT) of pulmonary tumors. , 2007, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[41]  I. Kawrakow Improved modeling of multiple scattering in the Voxel Monte Carlo model. , 1997, Medical physics.

[42]  C C Ling,et al.  Clinical experience with intensity modulated radiation therapy (IMRT) in prostate cancer. , 2000, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[43]  Samuel R Mazin,et al.  Emission-guided radiation therapy: biologic targeting and adaptive treatment. , 2010, Journal of the American College of Radiology : JACR.

[44]  Jian Z. Wang,et al.  Dosimetric advantages of IMRT simultaneous integrated boost for high-risk prostate cancer. , 2005, International journal of radiation oncology, biology, physics.

[45]  J O Deasy,et al.  Tomotherapy: a new concept for the delivery of dynamic conformal radiotherapy. , 1993, Medical physics.

[46]  Koichi Yamazaki,et al.  Feasibility of insertion/implantation of 2.0-mm-diameter gold internal fiducial markers for precise setup and real-time tumor tracking in radiotherapy. , 2003, International journal of radiation oncology, biology, physics.

[47]  Lei Zhu,et al.  Search for IMRT inverse plans with piecewise constant fluence maps using compressed sensing techniques. , 2009, Medical physics.

[48]  C. Langer,et al.  Concurrent chemoradiation using paclitaxel and carboplatin in locally advanced non-small cell lung cancer. , 1999, Seminars in radiation oncology.

[49]  S. Korreman,et al.  IGRT of prostate cancer; is the margin reduction gained from daily IG time-dependent? , 2006, Acta oncologica.

[50]  Mark P. Carol,et al.  Peacock™: A system for planning and rotational delivery of intensity‐modulated fields , 1995, Int. J. Imaging Syst. Technol..

[51]  J. Ciezki,et al.  Fluoroscopic study of tumor motion due to breathing: facilitating precise radiation therapy for lung cancer patients. , 2001, Medical physics.

[52]  Lei Zhu,et al.  TU‐G‐BRA‐04: Emission Guided Radiation Therapy: A Simulation Study of Lung Cancer Treatment with Automatic Tumor Tracking Using a 4D Digital Patient Model , 2012 .

[53]  R. Emery,et al.  Clinical experience using respiratory gated radiation therapy: comparison of free-breathing and breath-hold techniques. , 2004, International journal of radiation oncology, biology, physics.

[54]  Zhifei Wen,et al.  Development of a frameless stereotactic radiosurgery system based on real-time 6D position monitoring and adaptive head motion compensation , 2010, Physics in medicine and biology.

[55]  Dirk De Ruysscher,et al.  PET scans in radiotherapy planning of lung cancer. , 2010, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[56]  Steve B. Jiang,et al.  An experimental investigation on intra-fractional organ motion effects in lung IMRT treatments. , 2003, Physics in medicine and biology.

[57]  Timothy Solberg,et al.  Multi-institutional clinical experience with the Calypso System in localization and continuous, real-time monitoring of the prostate gland during external radiotherapy. , 2007, International journal of radiation oncology, biology, physics.

[58]  Radhe Mohan,et al.  Four-dimensional radiotherapy planning for DMLC-based respiratory motion tracking. , 2005, Medical physics.

[59]  J. Adler,et al.  Robotic Motion Compensation for Respiratory Movement during Radiosurgery , 2000, Computer aided surgery : official journal of the International Society for Computer Aided Surgery.