A Sub-10 nA DC-Balanced Adaptive Stimulator IC With Multi-Modal Sensor for Compact Electro-Acupuncture Stimulation

A compact electro-acupuncture (EA) system is proposed for a multi-modal feedback EA treatment. It is composed of a needle, a compact EA patch, and an interconnecting conductive thread. The 3 cm diameter compact EA patch is implemented with an adaptive stimulator IC and a small coin battery on the planar-fashionable circuit board (P-FCB) technology. The adaptive stimulator IC can form a closed current loop for even a single needle, and measure the electromyography (EMG) and the skin temperature to analyze the stimulation status as well as supply programmable stimulation current (40 μA-1 mA) with 5 different modes. The large time constant (LTC) sample and hold (S/H) current matching technique achieves the high-precision charge balancing ( <;10 nA) for the patient's safety. The measured data can be wirelessly transmitted to the external EA analyzer through the body channel communication (BCC) transceiver for the low power consumption. The external EA analyzer can show the patient's status, such as the muscle fatigue and the change of the skin temperature. Based on these analyses, the practitioner can adaptively change the stimulation parameters for the optimal treatment value. A 12.5 mm2 0.13 μm RF CMOS stimulator chip consumes 6.8 mW at 1.2 V supporting 32 different current levels. The proposed compact EA system is fully implemented and tested on the human body.

[1]  Maurits Ortmanns,et al.  A Neural Stimulator Frontend With High-Voltage Compliance and Programmable Pulse Shape for Epiretinal Implants , 2012, IEEE Journal of Solid-State Circuits.

[2]  Ji-Sheng Han,et al.  Electroacupuncture in the Treatment of Obesity , 2008, Neurochemical Research.

[3]  Hoi-Jun Yoo,et al.  A 1.12mW Continuous Healthcare Monitor Chip Integrated on a Planar Fashionable Circuit Board , 2008, 2008 IEEE International Solid-State Circuits Conference - Digest of Technical Papers.

[4]  Jisheng Han,et al.  Electroacupuncture: mechanisms and clinical application , 1998, Biological Psychiatry.

[5]  Maurits Ortmanns,et al.  An Active Approach for Charge Balancing in Functional Electrical Stimulation , 2010, IEEE Transactions on Biomedical Circuits and Systems.

[6]  Hoi-Jun Yoo,et al.  A sub-10nA DC-balanced adaptive stimulator IC with multimodal sensor for compact electro-acupuncture system , 2012, 2012 IEEE International Solid-State Circuits Conference.

[7]  M. Ortmanns,et al.  A 232-Channel Epiretinal Stimulator ASIC , 2007, IEEE Journal of Solid-State Circuits.

[8]  L. Straker,et al.  EMG median frequency changes in the neck-shoulder stabilizers of symptomatic office workers when challenged by different physical stressors. , 2005, Journal of electromyography and kinesiology : official journal of the International Society of Electrophysiological Kinesiology.

[9]  Andreas Demosthenous,et al.  An Integrated Implantable Stimulator That is Fail-Safe Without Off-Chip Blocking-Capacitors , 2008, IEEE Transactions on Biomedical Circuits and Systems.

[10]  B. Pomeranz,et al.  Monoaminergic mechanism of electroacupuncture analgesia , 1981, Brain Research.

[11]  B. Pomeranz,et al.  Electroacupuncture analgesia is mediated by stereospecific opiate receptors and is reversed by antagonists of type I receptors. , 1980, Life sciences.

[12]  Gerard O'Regan Texas Instruments , 1964, Nature.

[13]  Hoi-Jun Yoo,et al.  A 0.9V 2.6mW Body-Coupled Scalable PHY Transceiver for Body Sensor Applications , 2007, 2007 IEEE International Solid-State Circuits Conference. Digest of Technical Papers.

[14]  Allen McDaniels Electroacupuncture A Practical Manual and Resource , 2008 .

[15]  Dr. med. Gabriel Stux,et al.  Basics of Acupuncture , 1988, Springer Berlin Heidelberg.

[16]  J. Weiland,et al.  A variable range bi-phasic current stimulus driver circuitry for an implantable retinal prosthetic device , 2005, IEEE Journal of Solid-State Circuits.

[17]  Rahul Sarpeshkar,et al.  A Low-Power Blocking-Capacitor-Free Charge-Balanced Electrode-Stimulator Chip With Less Than 6 nA DC Error for 1-mA Full-Scale Stimulation , 2007, IEEE Transactions on Biomedical Circuits and Systems.

[18]  Nigel H. Lovell,et al.  CMOS neurostimulation ASIC with 100 channels, scaleable output, and bidirectional radio-frequency telemetry , 2001, IEEE Transactions on Biomedical Engineering.

[19]  S. Rendig,et al.  Reversal of reflex-induced myocardial ischemia by median nerve stimulation: a feline model of electroacupuncture. , 1998, Circulation.

[20]  Seulki Lee,et al.  A Wirelessly Powered Electro-Acupuncture Based on Adaptive Pulsewidth Monophase Stimulation , 2011, IEEE Transactions on Biomedical Circuits and Systems.

[21]  Ü. Tan,et al.  ELECTROACUPUNCTURE TREATMENT OF OBESITY WITH PSYCHOLOGICAL SYMPTOMS , 2007, The International journal of neuroscience.

[22]  J. Thomas Mortimer,et al.  Imbalanced biphasic electrical stimulation: Muscle tissue damage , 2006, Annals of Biomedical Engineering.

[23]  Seulki Lee,et al.  Planar Fashionable Circuit Board Technology and Its Applications , 2009 .

[24]  Hoi-Jun Yoo,et al.  Low-power NoC for high-performance SoC design , 2008 .

[25]  M. Sarek,et al.  EAV and Gemmotherapy - Medicine for the Next Millennium? (technique as a means to link eastern and western medicine) , 2005, 2005 IEEE Engineering in Medicine and Biology 27th Annual Conference.

[26]  Ya Tu,et al.  A Novel Uni-Acupoint Electroacupuncture Stimulation Method for Pain Relief , 2011, Evidence-based complementary and alternative medicine : eCAM.

[27]  Colin Lewis,et al.  Electroacupuncture – a Practical Manual and Resource , 2007 .