[Comparison of basic imaging properties for digital radiographic systems based on a direct conversion type of flat panel detector].

The purpose of this study was to evaluate the basic imaging properties of the two latest digital radiographic X-ray systems, namely, the DynaDirect Winscope 6000 (abbr. as System A, Toshiba Medical Systems) and the Sonialvision Safire Multi (abbr. as System B, Shimadzu Corp.). These systems were based on a direct-conversion flat panel detector (FPD) of amorphous selenium. The basic imaging properties of the two systems were evaluated by measuring characteristic curves, presampled modulation transfer functions (MTFs), and noise power spectra (NPS) using DICOM images to which no resampling was performed with a matrix size of 2048x2048. In addition, noise equivalent quanta (NEQ) and detective quantum efficiency (DQE) calculated from the result of the basic imaging properties were evaluated. The characteristic curves of the two systems showed quite high linearity. The MTFs of the two systems indicated high-resolution properties, as is well known to be an advantage of the direct conversion FPD system. However, the NPS of System A showed better performance than System B under the same exposures. Therefore, the DQE of System A was higher than that of System B at all spatial frequencies.

[1]  M L Giger,et al.  Investigation of basic imaging properties in digital radiography. 7. Noise Wiener spectra of II-TV digital imaging systems. , 1986, Medical physics.

[2]  東田善治 デジタル時代の医用画像情報技術セミナー V イメージ評価編 2 8. デジタル特性曲線の実践的測定法 , 2003 .

[3]  Kunio Doi,et al.  A simple method for determining the modulation transfer function in digital radiography , 1992, IEEE Trans. Medical Imaging.

[4]  M. Giger,et al.  Investigation of basic imaging properties in digital radiography. 2. Noise Wiener spectrum. , 1984, Medical physics.

[5]  H Fujita,et al.  Basic imaging properties of a computed radiographic system with photostimulable phosphors. , 1989, Medical physics.

[6]  清水正己 透視撮影におけるFPDシステムの進展2. Partire (日立メディコ) FPD搭載IVRデジタルイメージングシステム , 2003 .

[7]  Ehsan Samei,et al.  An experimental comparison of detector performance for direct and indirect digital radiography systems. , 2003, Medical physics.

[8]  Ehsan Samei Image quality in two phosphor-based flat panel digital radiographic detectors. , 2003, Medical physics.

[9]  鈴木英文 透視撮影におけるFPDシステムの進展7. 動画 (直接型) システム (島津製作所) 島津直接型FPDの有用性と搭載製品の展開 , 2003 .

[10]  Yoshiharu Higashida,et al.  [Image quality and detection performance of a direct digital radiography system]. , 2006, Nihon Hoshasen Gijutsu Gakkai zasshi.

[11]  島田和実 透視撮影におけるFPDシステムの進展4. Integris Allura9+ダイナミックフラットディテクタ (フィリップスメディカルシステムズ) , 2003 .

[12]  加野亜紀子 Medical Image Processing & Analysis Technology デジタル時代の医用画像情報技術セミナー V イメージ評価編 2 4.DQEの基礎と応用 , 2003 .

[13]  Takashi Maeda,et al.  [Effect on image data resampling in evaluation of the basic imaging properties for a digital radiographic system based on a flat panel detector]. , 2007, Nihon Hoshasen Gijutsu Gakkai zasshi.