Shallow Convolutional Neural Network for 3D Gamma Ray Localization in High Resolution PET

PET imaging inherently suffers from relatively poor spatial resolution compared with other clinical imaging modalities due to the size of the detector array elements, namely the scintillator crystals and readout pixels. Using 4-to-1 scintillator-to-readout coupling schemes, single-ended readout detector modules have been developed that simultaneously achieve depth-encoding readout to mitigate detector-based artifacts (i.e., parallax error) as well as sub-pixel spatial resolution equal to the scintillator crystal dimensions. Anger Logic-based localization schemes, such as energy weighted averaging, are commonly used to reconstruct the positions of photoelectric interaction sites within the modules, but these methods aren’t robust to edge and corner performance degradation, resulting in non-uniform position-dependent spatial performance. We explore the use of convolutional neural networks to perform 3D gamma ray localization with single-ended readout depth-encoding modules using Monte Carlo simulated PET data. Our approach to localizing gamma ray interactions in PET detectors outperforms conventional high resolution DOI readout methods, making it a step toward making cost-effective single-ended readout depth-encoding detector modules practically viable.

[1]  P. Dorenbos,et al.  Advances in Yield Calibration of Scintillators , 2008, IEEE Transactions on Nuclear Science.

[2]  G. Stringhini,et al.  A new method for depth of interaction determination in PET detectors , 2016, Physics in medicine and biology.

[3]  K. Kamada,et al.  Development of a Dual-Sided Readout DOI-PET Module Using Large-Area Monolithic MPPC-Arrays , 2013, IEEE Transactions on Nuclear Science.

[4]  S. Cherry,et al.  Using convolutional neural networks to estimate time-of-flight from PET detector waveforms , 2018, Physics in medicine and biology.

[5]  Yoshua Bengio,et al.  Gradient-based learning applied to document recognition , 1998, Proc. IEEE.

[6]  Jimmy Ba,et al.  Adam: A Method for Stochastic Optimization , 2014, ICLR.

[7]  G. Stringhini,et al.  Development of high-resolution detector module with depth of interaction identification for positron emission tomography , 2017 .

[8]  G. Stringhini,et al.  Development of a high resolution module for PET scanners , 2016 .

[9]  W. Moses Fundamental Limits of Spatial Resolution in PET. , 2011, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment.

[10]  M. G. Bisogni,et al.  Depth of interaction determination in monolithic scintillator with double side SiPM readout , 2017, EJNMMI Physics.

[11]  Peter Bruyndonckx,et al.  Monolithic scintillator PET detectors with intrinsic depth-of-interaction correction , 2009, Physics in medicine and biology.

[12]  Edoardo Charbon,et al.  System Tradeoffs in Gamma-Ray Detection Utilizing SPAD Arrays and Scintillators , 2010, IEEE Transactions on Nuclear Science.

[13]  G. Stringhini,et al.  On light sharing TOF-PET modules with depth of interaction and 157 ps FWHM coincidence time resolution , 2019, Physics in medicine and biology.