Experimental comparison of lesion detectability for four fully-3D PET reconstruction schemes

The objective of this work was to evaluate the lesion detectability performance of four fully-3D PET reconstruction schemes using experimentally acquired data. A multicompartment anthropomorphic phantom was setup to mimic whole-body FDG cancer imaging and scanned twelve times in 3D mode. Eight of the scans had twenty-six 68Ge "shell-less" lesions (6, 8, 10, 12, 16 mm diam.) placed throughout the phantom with various target:background ratios. This provided lesion-present and lesion-absent datasets with known truth appropriate for evaluating lesion detectability by localization receiver operating characteristics (LROC) methods. Four reconstruction schemes were studied: (1) Fourier rebinning (FORE) followed by 2D attenuation-weighted ordered-subsets expectation-maximization; (2) fully-3D AW-OSEM; (3) fully-3D ordinary-Poisson line-of- response (LOR-)OSEM; and (4) fully-3D LOR-OSEM with an accurate point-spread function (PSF) model. Two forms of LROC analysis were performed. First, a channelized non-prewhitened (CNPW) observer was used to optimize processing parameters (number of iterations, post-reconstruction filter) for the human observer study. Human observers then rated each image and selected the most-likely lesion location. The area under the LROC curve (ALROC) and the probability of correct localization were used as figures-of-merit. The results of the human observer study found no statistically significant difference between FORE and AW-OSEM3D (ALROC=0.41 and 0.36, respectively), an increase in lesion detection performance for LOR-OSEM3D (ALROC=0.45, p=0.076), and additional improvement with the use of the PSF model (ALROC=0.55, p=0.024). The numerical CNPW observer provided the same rankings in performance among algorithms, but obtained different values of ALROC. These results show improved lesion detection performance for the reconstruction algorithms with more sophisticated statistical and imaging models as compared to the previous-generation algorithms.

[1]  Dan J Kadrmas,et al.  LOR-OSEM: statistical PET reconstruction from raw line-of-response histograms , 2004, Physics in medicine and biology.

[2]  M. King,et al.  Human-observer receiver-operating-characteristic evaluation of attenuation, scatter, and resolution compensation strategies for (99m)Tc myocardial perfusion imaging. , 2003, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[3]  H.C. Gifford,et al.  Evaluation of multiclass model observers in PET LROC studies , 2004, IEEE Symposium Conference Record Nuclear Science 2004..

[4]  M. Chen,et al.  Randoms Mean Value Estimation with Exact Method for Ring PET Scanner , 2006, 2006 IEEE Nuclear Science Symposium Conference Record.

[5]  Thomas K. Lewellen,et al.  Modeling and incorporation of system response functions in 3-D whole body PET , 2006, IEEE Transactions on Medical Imaging.

[6]  T. Hebert,et al.  Fast methods for including attenuation in the EM algorithm , 1990 .

[7]  C. C. Watson New, faster, image-based scatter correction for 3D PET , 1999 .

[8]  Xuan Liu,et al.  Comparison of 3-D reconstruction with 3D-OSEM and with FORE+OSEM for PET , 2001, IEEE Transactions on Medical Imaging.

[9]  R. Swensson Unified measurement of observer performance in detecting and localizing target objects on images. , 1996, Medical physics.

[10]  W. W. Moses,et al.  List-mode maximum-likelihood reconstruction applied to positron emission mammography (PEM) with irregular sampling , 2000, IEEE Transactions on Medical Imaging.

[11]  Charles L. Byrne,et al.  Likelihood maximization for list-mode emission tomographic image reconstruction , 2001, IEEE Transactions on Medical Imaging.

[12]  Vladimir Y. Panin,et al.  Fully 3-D PET reconstruction with system matrix derived from point source measurements , 2006, IEEE Transactions on Medical Imaging.

[13]  Dan J. Kadrmas,et al.  Modeling the spatially-variant point spread function in a fast projector for improved fully-3D PET reconstruction , 2007 .

[14]  M.E. Casey,et al.  A /sup 68/Ge PET hot-sphere phantom with no cold shells , 2005, IEEE Nuclear Science Symposium Conference Record, 2005.

[15]  T. J. Hebert,et al.  Numerical evaluation of methods for computing tomographic projections , 1994 .

[16]  J. Hanley,et al.  The meaning and use of the area under a receiver operating characteristic (ROC) curve. , 1982, Radiology.

[17]  Michel Defrise,et al.  Exact and approximate rebinning algorithms for 3-D PET data , 1997, IEEE Transactions on Medical Imaging.

[18]  D. Chakraborty,et al.  Free-response methodology: alternate analysis and a new observer-performance experiment. , 1990, Radiology.

[19]  Dan J Kadrmas,et al.  Comparative evaluation of lesion detectability for 6 PET imaging platforms using a highly reproducible whole-body phantom with (22)Na lesions and localization ROC analysis. , 2002, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[20]  L. Shepp,et al.  Maximum Likelihood Reconstruction for Emission Tomography , 1983, IEEE Transactions on Medical Imaging.

[21]  Michael A. King,et al.  Assessment of scatter compensation strategies for /sup 67/Ga tumor SPECT using numerical observers and human LROC studies , 2002, 2002 IEEE Nuclear Science Symposium Conference Record.

[22]  Eric C. Frey,et al.  Application of task-based measures of image quality to optimization and evaluation of three-dimensional reconstruction-based compensation methods in myocardial perfusion SPECT , 2002, IEEE Transactions on Medical Imaging.

[23]  H H Barrett,et al.  List-mode likelihood. , 1997, Journal of the Optical Society of America. A, Optics, image science, and vision.

[24]  Stefan Eberl,et al.  Optimizing injected dose in clinical PET by accurately modeling the counting-rate response functions specific to individual patient scans. , 2005, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[25]  T. M. Peters Algorithms for Fast Back- and Re-Projection in Computed Tomography , 1981 .

[26]  P F Judy,et al.  Evaluation of video gray-scale display. , 1992, Medical physics.

[27]  Dan J. Kadrmas,et al.  Rotate-and-Slant Projector for Fast LOR-Based Fully-3-D Iterative PET Reconstruction , 2008, IEEE Transactions on Medical Imaging.

[28]  D. Townsend,et al.  Physical Performance and Clinical Workflow of a new LSO HI-REZ PET/CT Scanner , 2006, 2006 IEEE Nuclear Science Symposium Conference Record.

[29]  M. Defrise,et al.  A fast rebinning algorithm for 3D positron emission tomography using John's equation , 1999 .

[30]  C. Comtat,et al.  Fast reconstruction of 3-D PET data with accurate statistical modeling , 1997 .

[31]  K. Lange,et al.  EM reconstruction algorithms for emission and transmission tomography. , 1984, Journal of computer assisted tomography.

[32]  Eric C Frey,et al.  Assessment of scatter compensation strategies for (67)Ga SPECT using numerical observers and human LROC studies. , 2004, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[33]  H. Malcolm Hudson,et al.  Accelerated image reconstruction using ordered subsets of projection data , 1994, IEEE Trans. Medical Imaging.

[34]  Michael A. King,et al.  LROC analysis of detector-response compensation in SPECT , 2000, IEEE Transactions on Medical Imaging.

[35]  Michael A. King,et al.  Single-slice versus multi-slice display for human-observer lesion-detection studies , 1999 .

[36]  M. Defrise,et al.  Fast reconstruction of 3-D PET data with accurate statistical modeling , 1997, 1997 IEEE Nuclear Science Symposium Conference Record.

[37]  P. Joseph An Improved Algorithm for Reprojecting Rays through Pixel Images , 1983, IEEE Transactions on Medical Imaging.