Experimental validation tests of fast Fourier transform convolution and multigrid superposition algorithms for dose calculation in low-density media.

BACKGROUND AND PURPOSE Modern conformal radiotherapy treatments require accurate dose calculation in any relevant clinical situation. One of these situations is the treatment of lung tumors, where irradiation has to be planned under challenging conditions for dose calculation. In this study we assess the errors in dose values predicted by fast Fourier transform convolution (FFTC) and multigrid superposition (MGS) algorithms implemented in a commercial treatment planning system (TPS). MATERIALS AND METHODS FFTC and MGS algorithms were used in a FOCUS 3.0.0 (Computerized Medical Systems, USA) to calculate doses in treatment plans using photon beams of 6 and 25 MV nominal energy from a Saturne 43 linac (GE Medical Systems, USA). A 10x10-cm beam irradiating a mediastinum-lung and a thoracic wall-lung-thoracic wall modeled geometry was assessed. The calculated data were compared with measurements performed with radiographic films and ionization chamber. RESULTS FFTC algorithm leads to an average deviation from ionometric dose measurements of over 10%. Discrepancies between measured and calculated beam fringe values (distance between 50 and 90% isodose lines) of up to 8 mm were observed. For MGS algorithm, all the points assessed in both geometries fulfilled the 3%-3 mm accuracy criteria and the average deviation of absolute dose was about 1%. A maximum of 3 mm deviation in the beam fringe for any depth was found and was within 2 mm beyond the buildup region. Deviations between ionometric and film measurements were within 3%. CONCLUSIONS MGS algorithm assesses with reasonable accuracy dose distributions and absolute dose in inhomogeneous regions like the lung region. Therefore, and respecting the inhomogeneity dose calculation, the system could be used in routine clinical practice and in dose-escalation programs. This is not true in the case of FFTC algorithm which leads to errors greater than 10% in the absolute dose calculation and underestimates the beam fringe by up to 8 mm.

[1]  A. Ahnesjö Collapsed cone convolution of radiant energy for photon dose calculation in heterogeneous media. , 1989, Medical physics.

[2]  G. Desobry,et al.  Effect of low-density lateral interfaces on soft-tissue doses. , 1997, International journal of radiation oncology, biology, physics.

[3]  G. Starkschall,et al.  American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: quality assurance for clinical radiotherapy treatment planning. , 1998, Medical physics.

[4]  M Miften,et al.  Implementation of FFT convolution and multigrid superposition models in the FOCUS RTP system. , 2000, Physics in medicine and biology.

[5]  M Engelsman,et al.  Impact of simple tissue inhomogeneity correction algorithms on conformal radiotherapy of lung tumours. , 2001, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[6]  A. Kapur,et al.  Comparison of RTP dose distributions in heterogeneous phantoms with the beam Monte Carlo simulation system , 2001, Journal of applied clinical medical physics.

[7]  J. Battista,et al.  A convolution method of calculating dose for 15-MV x rays. , 1985, Medical physics.

[8]  L. Martínez,et al.  Un método para la conversión del número Hounsfield en densidad electrónica y para la obtención de la energía efectiva en los escáneres CT , 2002 .

[9]  J C Rosenwald,et al.  Variation of sensitometric curves of radiographic films in high energy photon beams. , 2001, Medical physics.

[10]  A. Ahnesjö,et al.  Dose calculations for external photon beams in radiotherapy. , 1999, Physics in medicine and biology.

[11]  R. Cloutier Tissue Substitutes in Radiation Dosimetry and Measurement. , 1989 .

[12]  J J Battista,et al.  Generation of photon energy deposition kernels using the EGS Monte Carlo code. , 1988, Physics in medicine and biology.

[13]  P Keall,et al.  Accounting for primary electron scatter in x-ray beam convolution calculations. , 1995, Medical physics.

[14]  J. Cunningham,et al.  Evaluation of lung dose correction methods for photon irradiations of thorax phantoms. , 1989, International journal of radiation oncology, biology, physics.

[15]  I. Rosen,et al.  Beam‐commissioning methodology for a three‐dimensional convolution/superposition photon dose algorithm , 2000, Journal of applied clinical medical physics.

[16]  R. Mohan,et al.  The impact of electron transport on the accuracy of computed dose. , 2000, Medical physics.

[17]  B Mijnheer,et al.  Tolerances for the accuracy of photon beam dose calculations of treatment planning systems. , 2001, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[18]  K. Ekstrand,et al.  Pitfalls in the use of high energy X rays to treat tumors in the lung. , 1990, International journal of radiation oncology, biology, physics.

[19]  M E Young,et al.  Changes in the dose-profile of a 10 MV x-ray beam within and beyond low density material. , 1982, Medical physics.