A comparison of a similarity-based and a feature-based 2-D-3-D registration method for neurointerventional use

Two-dimensional (2-D)-to-three-dimensional (3-D) registration can improve visualization which may aid minimally invasive neurointerventions. Using clinical and phantom studies, two state-of-the-art approaches to rigid registration are compared quantitatively: an intensity-based algorithm using the gradient difference similarity measure; and an iterative closest point (ICP)-based algorithm. The gradient difference approach was found to be more accurate, with an average registration accuracy of 1.7 mm for clinical data, compared to the ICP-based algorithm with an average accuracy of 2.8 mm. In phantom studies, the ICP-based algorithm proved more reliable, but with more complicated clinical data, the gradient difference algorithm was more robust. Average computation time for the ICP-based algorithm was 20 s per registration, compared with 14 min and 50 s for the gradient difference algorithm.

[1]  Alan Liu,et al.  3D/2D Registration via Skeletal Near Projective Invariance in Tubular Objects , 1998, MICCAI.

[2]  Jacques A. de Guise,et al.  A method for modeling noise in medical images , 2004, IEEE Transactions on Medical Imaging.

[3]  L. Joskowicz,et al.  Gradient-based 2-D/3-D rigid registration of fluoroscopic X-ray to CT , 2003, IEEE Transactions on Medical Imaging.

[4]  Peter Kazanzides,et al.  Anatomy-based registration of CT-scan and intraoperative X-ray images for guiding a surgical robot , 1998, IEEE Transactions on Medical Imaging.

[5]  O Levrier,et al.  Endovascular treatment of intracranial aneurysms. , 1996, AJNR. American journal of neuroradiology.

[6]  Erhard Klotz,et al.  Real-Time Distortion Correction Of Digital X-Ray II/TV Systems: An Application Example , 1989, Other Conferences.

[7]  Søren Ingvor Olsen Noise Variance Estimation in Images , 1993 .

[8]  Robert A. McLaughlina,et al.  Intensity-based Registration versus Feature-based Registration for Neurointerventions , 2001 .

[9]  David W. Murray,et al.  Visual head tracking and slaving for visual telepresence , 1996, Proceedings of IEEE International Conference on Robotics and Automation.

[10]  Y Trousset,et al.  Intracranial aneurysms: clinical value of 3D digital subtraction angiography in the therapeutic decision and endovascular treatment. , 2001, Radiology.

[11]  Paul J. Besl,et al.  A Method for Registration of 3-D Shapes , 1992, IEEE Trans. Pattern Anal. Mach. Intell..

[12]  J. Alison Noble,et al.  Fusing speed and phase information for vascular segmentation of phase contrast MR angiograms , 2002, Medical Image Anal..

[13]  David J. Hawkes,et al.  2D-3D Intensity Based Registration of DSA and MRA - A Comparison of Similarity Measures , 2002, MICCAI.

[14]  B Likar,et al.  Automatic extraction of corresponding points for the registration of medical images. , 1999, Medical physics.

[15]  Tomaz Slivnik,et al.  3-D/2-D registration of CT and MR to X-ray images , 2003, IEEE Transactions on Medical Imaging.

[16]  Marie-Odile Berger,et al.  Fully Automatic 3D/2D Subtracted Angiography Registration , 1999, MICCAI.

[17]  David L. Wilson,et al.  Comparison of algorithms for combining X-ray angiography images , 2001, IEEE Transactions on Medical Imaging.

[18]  J. Alison Noble,et al.  Statistical 3D Vessel Segmentation Using a Rician Distribution , 1999, MICCAI.

[19]  D. R. Fish,et al.  A patient-to-computed-tomography image registration method based on digitally reconstructed radiographs. , 1994, Medical physics.

[20]  Kazuyuki Kojima,et al.  Clinical benefits of rotational 3D angiography in endovascular treatment of ruptured cerebral aneurysm. , 2002, AJNR. American journal of neuroradiology.

[21]  Hiroshi Iseki,et al.  Interactive virtualized display system for intravascular neurosurgery , 1997, CVRMed.

[22]  S.A.M. Baert,et al.  Three-dimensional guide-wire reconstruction from biplane image sequences for integrated display in 3-D vasculature , 2003, IEEE Transactions on Medical Imaging.

[23]  Kawal S. Rhode,et al.  Intensity-based 2-D - 3-D registration of cerebral angiograms , 2003, IEEE Transactions on Medical Imaging.

[24]  Nicholas Ayache,et al.  3D-2D Projective Registration of Free-Form Curves and Surfaces , 1997, Comput. Vis. Image Underst..

[25]  David J. Hawkes,et al.  A Comparison of 2D-3D Intensity-Based Registration and Feature-Based Registration for Neurointerventions , 2002, MICCAI.

[26]  D L Hill,et al.  Validation of a two- to three-dimensional registration algorithm for aligning preoperative CT images and intraoperative fluoroscopy images. , 2001, Medical physics.

[27]  Robert C. Bolles,et al.  Random sample consensus: a paradigm for model fitting with applications to image analysis and automated cartography , 1981, CACM.

[28]  J. Alison Noble,et al.  Real-Time Registration of 3D Cerebral Vessels to X-ray Angiograms , 1998, MICCAI.

[29]  S. Rudin,et al.  Accurate characterization of image intensifier distortion. , 1991, Medical physics.

[30]  Alejandro F. Frangi,et al.  Guide Wire Tracking During Endovascular Interventions , 2000, International Conference on Medical Image Computing and Computer-Assisted Intervention.

[31]  Max A. Viergever,et al.  CTA-based angle selection for diagnostic and interventional angiography of saccular intracranial aneurysms , 1998, IEEE Transactions on Medical Imaging.

[32]  J. Alison Noble,et al.  Determining X-ray projections for coil treatments of intracranial aneurysms , 1999, IEEE Transactions on Medical Imaging.

[33]  Jürgen Weese,et al.  A comparison of similarity measures for use in 2-D-3-D medical image registration , 1998, IEEE Transactions on Medical Imaging.

[34]  M. Murphy An automatic six-degree-of-freedom image registration algorithm for image-guided frameless stereotaxic radiosurgery. , 1997, Medical physics.

[35]  Graeme Penney Registration of tomographic images to X-ray projections for use in image guided interventions , 2000 .