Performance Limitations Arising in Closed Loop Control of Blood Glucose in Type 1 Diabetes

Abstract This paper presents a preliminary study of performance limitations that arise in the closed-loop control of blood glucose, using an autonomous artificial pancreas. It is shown that a major source of limitations is due to model uncertainty, specifically due to the combined effect of the insulin infusion system (IIS), the continuous glucose monitor (CGM) and the human glucose regulatory system. It is argued that the uncertainty associated with each of these elements compromises the achievable closed-loop bandwidth, and, in the presence of disturbances, e.g. meal intake and exercise, the closed-loop response will necessarily be poor. A proposition to overcome this problem is given based on feedforward action.

[1]  Marc D. Breton,et al.  Modular Closed-Loop Control of Diabetes , 2012, IEEE Transactions on Biomedical Engineering.

[2]  C. Cobelli,et al.  Progress in Development of an Artificial Pancreas , 2009, Journal of diabetes science and technology.

[3]  Janet M. Allen,et al.  Day and Night Closed-Loop Control in Adults With Type 1 Diabetes , 2013, Diabetes Care.

[4]  R. Hovorka Continuous glucose monitoring and closed‐loop systems , 2006, Diabetic medicine : a journal of the British Diabetic Association.

[5]  J. Doyle,et al.  Essentials of Robust Control , 1997 .

[6]  F. Doyle,et al.  Quest for the Artificial Pancreas: Combining Technology with Treatment , 2010, IEEE Engineering in Medicine and Biology Magazine.

[7]  Tarunraj Singh,et al.  Blood glucose control algorithms for type 1 diabetic patients: A methodological review , 2013, Biomed. Signal Process. Control..

[8]  Garry M. Steil,et al.  Identification of Intraday Metabolic Profiles during Closed-Loop Glucose Control in Individuals with Type 1 Diabetes , 2009, Journal of diabetes science and technology.

[9]  Giuseppe De Nicolao,et al.  MPC based Artificial Pancreas: Strategies for individualization and meal compensation , 2012, Annu. Rev. Control..

[10]  G. Steil,et al.  Closed-loop insulin delivery-the path to physiological glucose control. , 2004, Advanced drug delivery reviews.

[11]  A H Clemens,et al.  Closed-loop and Open-loop Devices for Blood Glucose Control in Normal and Diabetic Subjects , 1978, Diabetes.

[12]  Clare E. Collins,et al.  Influence of and Optimal Insulin Therapy for a Low–Glycemic Index Meal in Children With Type 1 Diabetes Receiving Intensive Insulin Therapy , 2008, Diabetes Care.

[13]  Graham C. Goodwin,et al.  Feedforward model predictive control , 2011, Annu. Rev. Control..

[14]  Graham C. Goodwin,et al.  Control System Design , 2000 .

[15]  B. Bequette A critical assessment of algorithms and challenges in the development of a closed-loop artificial pancreas. , 2005, Diabetes technology & therapeutics.

[16]  Graham C. Goodwin,et al.  Fundamental Limitations in Filtering and Control , 1997 .

[17]  Bruce R. King,et al.  Changes in Altitude Cause Unintended Insulin Delivery From Insulin Pumps , 2011, Diabetes Care.

[18]  Dario Iafusco,et al.  Timing of bolus in children with type 1 diabetes using continuous subcutaneous insulin infusion (TiBoDi Study). , 2010, Diabetes technology & therapeutics.

[19]  W. Schenk,et al.  Does Physiological Blood Glucose Control Require an Adaptive Control Strategy? , 1987, IEEE Transactions on Biomedical Engineering.

[20]  G. Steil,et al.  Closed-loop insulin delivery – what lies between where we are and where we are going? , 2005, Expert opinion on drug delivery.

[21]  R. Bergman Minimal Model: Perspective from 2005 , 2006, Hormone Research in Paediatrics.